Skip to content

RELIANCE LITe 48061000Recalls & reported problems

Stryker Corporation · T-HANDLE, QUICK CONNECT · Class 1

All models of RELIANCE LITe (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Reliance LITe T-handle is a handheld quick-connect tool used in orthopedic surgery. It has 24 adverse event reports, with no deaths and no injuries; reports mostly describe the handle or its attachments fracturing or jamming during surgery, sometimes leaving a fragment in the wound that was retrieved, with procedures typically completed without delay. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 07613327263671), received between Apr 26, 2016 and May 15, 2026 — classified as 24 malfunctions.

The most frequently reported specific device problem is “Fracture” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture20
  • Break13
  • Mechanical Jam3
  • Detachment Of Device Component2
  • Material Integrity Problem2
  • Mechanical Problem2
  • Material Fragmentation1

Patient problems

  • No Known Impact Or Consequence To Patient14
  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2026Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT THE SECURING MECHANISM OF A RELIANCE LITE T-HANDLE QUICK CONNECT FRACTURED INTRA-OPERATIVELY. IT IS CURRENTLY UNKNOWN IF THE FRACTURED COMPONENT CAME INTO CONTACT WITH THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25184911 ↗
  • MalfunctionFeb 24, 2026Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT THE HANDLE ATTACHMENTS OF TWO RELIANCE LITE T- HANDLES FRACTURED INTRA-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE FRACTURED COMPONENT ENTERED THE PATIENT WOUND SITE AND WAS SUBSEQUENTLY RETRIEVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT CAPTURES THE SECOND OF TWO HANDLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434250 ↗
  • MalfunctionFeb 24, 2026Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT THE HANDLE ATTACHMENTS OF TWO RELIANCE LITE T- HANDLES FRACTURED INTRA-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE FRACTURED COMPONENT ENTERED THE PATIENT WOUND SITE AND WAS SUBSEQUENTLY RETRIEVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT CAPTURES THE FIRST OF TWO HANDLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434247 ↗
  • MalfunctionSep 9, 2025Fracture; Fracture

    A CUSTOMER REPORTED THAT A FRAGMENT FROM A RELIANCE LITE T- HANDLE, QUICK CONNECT FRACTURED OFF INTRA-OPERATIVELY. THE FRAGMENT WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23001492 ↗
  • MalfunctionMay 15, 2025Fracture; Fracture

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031875 ↗
  • MalfunctionAug 1, 2022Fracture; Mechanical Problem; Fracture; Mechanical Problem

    IT WAS REPORTED THAT TWO (2) RELIANCE LITE T- HANDLES WERE UNABLE TO BE USED INTRA-OPERATIVELY; THEY NO LONGER CONNECT TO THE SHAVERS. THE SURGERY WAS COMPLETED SUCCESSFULLY. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED. THIS REPORT CAPTURES THE FIRST HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15139512 ↗
  • MalfunctionAug 1, 2022Mechanical Problem; Mechanical Jam; Mechanical Problem; Mechanical Jam

    IT WAS REPORTED THAT TWO RELIANCE LITE T- HANDLES WERE UNABLE TO BE USED INTRA-OPERATIVELY; THEY NO LONGER CONNECT TO THE SHAVERS. THE SURGERY WAS COMPLETED SUCCESSFULLY. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED. THIS REPORT CAPTURES THE SECOND HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15139479 ↗
  • MalfunctionJul 27, 2022Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT A RELIANCE LITE T HANDLE BROKE INTRA-OPERATIVELY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15115980 ↗
  • MalfunctionJun 1, 2022Fracture; Mechanical Jam; Fracture; Mechanical Jam

    IT WAS REPORTED THAT A RELIANCE T-HANDLE WAS JAMMING OUTSIDE THE OPERATING ROOM. THERE WAS NO ASSOCIATED PROCEDURE WITH THIS EVENT, THEREFORE, NO ADVERSE CONSEQUENCES TO ANY PATIENTS WERE SEEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14566795 ↗
  • MalfunctionMar 8, 2022Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT A RELIANCE T-HANDLE IS HAVING TROUBLE RELEASING SHAFTS OUTSIDE THE OR. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13697286 ↗
  • MalfunctionMar 21, 2018Fracture; Break; Fracture

    IT WAS REPORTED THE INNER SHAFT OF THE HANDLE BROKE WHILE DISTRACTING THE DISC SPACE. THE SURGERY WAS SUCCESSFULLY COMPLETED. THERE ARE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7358009 ↗
  • MalfunctionJan 29, 2018Fracture; Material Fragmentation; Fracture

    IT WAS REPORTED THAT THE QUICK RELEASE HANDLES BROKE DURING SURGERY. ALL THE BROKEN FRAGMENTS WERE RETRIEVED. NO ADVERSE HEALTH CONSEQUENCES WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7224298 ↗
  • MalfunctionNov 22, 2017Break; Fracture; Fracture

    IT WAS REPORTED THAT; WHILE USING THE SHAVERS IN THE DISC SPACE, THE INNER TABS ON BLUE T HANDLE BROKE ON BOTH T HANDLES IN THE RELIANCE LITE INSTRUMENTS SET.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7052228 ↗
  • MalfunctionNov 22, 2017Break; Fracture; Fracture

    IT WAS REPORTED THAT; WHILE USING THE SHAVERS IN THE DISC SPACE, THE INNER TABS ON BLUE T HANDLE BROKE ON BOTH T HANDLES IN THE RELIANCE LITE INSTRUMENTS SET

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7052221 ↗
  • MalfunctionNov 21, 2017Break; Fracture; Fracture

    IT WAS REPORTED THAT; DURING A LATERAL CASE, TWO SEPARATE TRIAL/SHAVER HANDLES BROKE, HALF THE QUICK RELEASE MECHANISM FELL ONTO THE FIELD.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7050037 ↗
  • MalfunctionJul 27, 2017Fracture; Material Integrity Problem; Break; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT; DOCTOR WAS TURNING A SHAVER IN A DISC SPACE WHEN A HANDLE BROKE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6746566 ↗
  • MalfunctionApr 17, 2017Break; Detachment Of Device Component; Fracture; Material Integrity Problem; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT; UPON IMPACTING THE HANDLE, A PIECE OF METAL BROKE OFF THE BOTTOM OF T HANDLE. THE PIECE WAS RETRIEVED IMMEDIATELY. THE SURGEON THEN USED THE SLAP HAMMER TO REMOVE THE TRAIL, STILL ATTACHED TO THE T HANDLE, AND UPON I,PARTING THE HAMMER ON THE HANDLE, THE BLUE T PART OF THE HANDLE REMOVED, STILL LEAVING THE TRIAL IN PLACE. THE SLAP HAMMER WAS REATTACHED TO THE TRIAL AND SUCCESSFULLY REMOVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6498756 ↗
  • MalfunctionNov 15, 2016Break

    IT WAS REPORTED THAT; HANDLE BROKE WHILE TRIALING

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6104750 ↗
  • MalfunctionNov 15, 2016Fracture; Break; Fracture

    IT WAS REPORTED THAT; HANDLE BROKE WHILE TRIALING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6104714 ↗
  • MalfunctionOct 10, 2016Break; Detachment Of Device Component

    A SHAVER TURNED WITHIN THE T HANDLE BREAKING A PART SOME OF THE INSIDE OF THE T HANDLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6015464 ↗
  • MalfunctionAug 24, 2016Break; Fracture; Fracture

    IT WAS REPORTED THAT THE HANDLE BROKE WHEN TAKE WAS REMOVING FROM TRIAL

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5901406 ↗
  • MalfunctionMay 25, 2016Break; Fracture; Fracture

    IT WAS REPORTED THAT THE SURGEON WAS USING T-HANDLE WITH TRIAL ATTACHMENT IN DISC SPACE, HE THEN BEGAN TO REMOVE THE TRIAL WHEN HE FELT SOMETHING "GIVE WAY" OR BREAK.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5679819 ↗
  • MalfunctionApr 26, 2016Break; Fracture; Fracture

    IT WAS REPORTED THAT TWO T HANDLES BROKE. DOCTOR WAS TRYING TO DISTRACT VERY TIGHT DISC SPACE AND THE TWO T HANDLES BROKE. USED ANOTHER T HANDLE FROM A BACKUP TRAY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5607704 ↗
  • MalfunctionApr 26, 2016Break; Fracture; Fracture

    IT WAS REPORTED THAT TWO T HANDLES BROKE. DOCTOR WAS TRYING TO DISTRACT VERY TIGHT DISC SPACE AND THE TWO T HANDLES BROKE. USED ANOTHER T HANDLE FROM A BACKUP TRAY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5607430 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page