AVS 48350951Recalls & reported problems
Stryker Corporation · PL INSERTER 8-9 STRAIGHT · Class 1
All models of AVS (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327267914), received between Mar 2, 2018 and Nov 21, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- Material Deformation3
- Difficult to Remove2
- Mechanical Jam2
- Break1
- Corroded1
- Deformation Due to Compressive Stress1
- Disconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Fracture; Fracture
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23618872 ↗MalfunctionSep 29, 2025Difficult to Remove; Mechanical Jam; Difficult to Remove; Mechanical Jam
A COMPANY REPRESENTATIVE REPORTED THAT TWO AVS PL SPACER INSERTERS WERE DIFFICULT TO DISASSEMBLE FROM THEIR IMPLANTS INTRA-OPERATIVELY AND THAT THE DEVICES APPEARED TO BE CROSS-THREADED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT CAPTURES THE FIRST OF TWO INSERTERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23176176 ↗MalfunctionSep 29, 2025Difficult to Remove; Mechanical Jam; Difficult to Remove; Mechanical Jam
A COMPANY REPRESENTATIVE REPORTED THAT TWO AVS PL SPACER INSERTERS WERE DIFFICULT TO DISASSEMBLE FROM THEIR IMPLANTS INTRA-OPERATIVELY AND THAT THE DEVICES APPEARED TO BE CROSS-THREADED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT CAPTURES THE SECOND OF TWO INSERTERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23176090 ↗MalfunctionJul 2, 2024Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT THE TIP OF AN AVS PL INSERTER FRACTURED OFF INTRA-OPERATIVELY DURING CAGE INSERTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19661260 ↗MalfunctionMar 15, 2024Corroded; Fracture; Corroded; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT THE INNER SHAFT OF AN AVS SPACER STRAIGHT INSERTER FRACTURED INTRA-OPERATIVELY WHILE THE SURGEON WAS REMOVING THE CAGE TO REPOSITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18913420 ↗MalfunctionOct 28, 2022Material Deformation; Material Deformation
A COMPANY REPRESENTATIVE REPORTED THAT AN AVS PL SPACER STRAIGHT INSERTER BENT INTRA-OPERATIVELY. IT WAS FURTHER REPORTED VIA FOLLOW UP COMMUNICATION THAT THE INSERTER WAS USED AT A DIFFICULT ANGLE AND EXCESSIVE FORCE WAS APPLIED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A REPLACEMENT INSERTER WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15693709 ↗MalfunctionAug 3, 2022Fracture; No Apparent Adverse Event; Fracture; No Apparent Adverse Event
IT WAS REPORTED THAT THE TIP OF AN AVS PL SPACER STRAIGHT INSERTER BROKE OFF INTRA-OPERATIVELY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15161129 ↗MalfunctionJun 30, 2022Deformation Due to Compressive Stress; Material Deformation; Deformation Due to Compressive Stress; Material Deformation
A COMPANY REPRESENTATIVE REPORTED THAT AN AVS PL INSERTER BENT AT THE DISTAL TIP AND MIDDLE DURING INTRA-OPERATIVE IMPLANT PLACEMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14875781 ↗MalfunctionMar 2, 2018Material Deformation; Break; Disconnection; Material Deformation
IT WAS REPORTED THAT UPON STARTING RATCHETING THE INSTRUMENT WASN'T DISCONNECTING FROM THE IMPLANT AND BROKE. TIGHT SPACE WAS OBSERVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND THERE ARE NO REPORTS OF ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7311386 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page