AVS NAVIGATOR 48394005Recalls & reported problems
Stryker Corporation · DYNAMIC DISTRACTOR WITH HANDLE · Class 1
All models of AVS NAVIGATOR (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327270648), received between Mar 7, 2017 and Feb 16, 2018 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break4
- Fracture3
- Detachment Of Device Component1
- Device Operates Differently Than Expected1
- Difficult to Open or Close1
- Noise, Audible1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2018Fracture; Device Operates Differently Than Expected; Noise, Audible; Fracture
THE SURGEON REPORTED THAT A DISTRACTOR FAILED SUDDENLY DURING DISTRACTION, DURING A TLIF MINI INVASIVE L4 L5 SURGERY. THE SURGEON REPORTED HEARING A "CLACK" AND THOUGHT A PIECE BROKE INSIDE THE MECHANISM. THE MECHANISM DID NOT SEIZE. THE SURGEON WAS ABLE TO RETRACT THE INSTRUMENT TO REMOVE IT FROM THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7278677 ↗MalfunctionDec 28, 2017Break; Detachment Of Device Component; Fracture; Fracture
IT WAS REPORTED THAT; HANDLE BROKE OFF OF INSTRUMENT AND PART OF IT IS STUCK IN THREAD.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7150884 ↗MalfunctionAug 28, 2017Break
IT WAS REPORTED THAT; THIS DEVICE WAS USED AT AN EXTREME ANGLE UNDER EXCEEDING PRESSURE. UPDATE 8/2/2017: THE TIP BROKE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 6825873 ↗MalfunctionAug 28, 2017Break
IT WAS REPORTED THAT: THIS DEVICE WAS USED AT AN EXTREME ANGLE UNDER EXCEEDING PRESSURE UPDATE (B)(6) 2017: THE TIP BROKE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 6825871 ↗MalfunctionMar 7, 2017Break; Fracture; Difficult to Open or Close; Fracture
IT WAS REPORTED THAT; CENTRAL STERILE SERVICES OF THE CLINIC, REPORTED THAT "THE DISTRACTOR COULD NOT OPEN COMPLETELY, THE HANDLE BROKE INSIDE BY UNSCREWING IT DURING A MEDICAL PROCEDURE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6384168 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page