Stryker Corporation · CONNECTOR HEX SOCKET FOR TORQUE WRENCH · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327273328), received between May 11, 2016 and Sep 5, 2018 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break5
- Fracture3
- Detachment Of Device Component1
- Material Integrity Problem1
- Sticking1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2018Break; Fracture; Fracture
IT WAS REPORTED THAT THE HEX SOCKET BROKE DURING FINAL TIGHTENING. ALL BROKEN FRAGMENTS WERE RETRIEVED AND THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE INSTRUMENT, WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7847928 ↗MalfunctionAug 14, 2018Break; Fracture; Fracture
IT IS REPORTED THAT THE TIP HEX SOCKET OF A TORQUE WRENCH BROKE OFF IN A CONNECTOR WHEN FINAL TIGHTENING. NO ADVERSE CONSEQUENCES TO THE PATIENT IS REPORTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7782883 ↗MalfunctionFeb 9, 2018Break; Detachment Of Device Component; Fracture; Fracture
WHILE PERFORMING FINAL TIGHTENING, THE 3.5 HEX SOCKET BROKE OFF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACKUP ATTACHMENT. THERE WERE NO ADVERSE HEALTH CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7259579 ↗MalfunctionMay 25, 2016Break
IT WAS REPORTED THAT THE HEX ATTACHMENT FOR THE TORQUE DRIVER BROKE WHILE TORQUING THE CONNECTORS, HAD ANOTHER HEX ATTACHMENT THAT WAS USED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5679620 ↗MalfunctionMay 11, 2016Break; Sticking; Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT; TORQUE WRENCH TIP BROKE WHILE FINAL TIGHTENING. TIP WAS THEN STUCK IN CONNECTOR AND HAD TO BURR IT OUT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5648149 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.