Skip to content

AVS 48350952LRecalls & reported problems

Stryker Corporation · PL SPACER INSERTER WITH LASER MARKING · Class 1

All models of AVS (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327276817), received between Oct 29, 2022 and Nov 21, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT A PATIENT WITH NEUROLOGICAL SYMPTOMS WAS REVISED APPROXIMATELY 2.5 YEARS POST-OPERATIVELY TO ADDRESS TWO MIGRATED TRITANIUM POSTERIOR LUMBAR CAGES (ADDRESSED IN PI (B)(4)). WHILE ATTEMPTING TO REMOVE THE TWO LUMBAR CAGES WITH AN AVS SPACER INSERTER, BOTH CAGES FRACTURED AND THE INSERTER DEFORMED INTRA-OPERATIVELY. THE FRACTURED COMPONENTS OF THE CAGES WERE EXPLANTED, HOWEVER THE REST OF THE CAGES REMAINED IMPLANTED. THE CAGE SURFACES WERE SMOOTHED AND DECOMPRESSION WAS SUCCESSFULLY COMPLETED TO ALLEVIATE THE NEUROLOGICAL SYMPTOMS. THIS REPORT CAPTURES THE INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20747613 ↗
  • MalfunctionOct 29, 2022Material Deformation; Material Deformation

    THIS RECORD IS A CONSOLIDATION OF Q3 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES ONE INSTANCE OF INTRAOPERATIVE AVS PL STRAIGHT INSERTER DEFORMATION. THE ASSOCIATED PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15695986 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page