T2 ALPHA 2337-1036SRecalls & reported problems
Stryker GmbH · Femoral Nail PF, RIGHT · Class 2 · Implant
All models of T2 ALPHA (123)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327289145), received between Aug 21, 2024 and Aug 21, 2024 — classified as 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "POST-MARKET CLINICAL EVALUATION OF THE TREATMENT OF FEMUR FRACTURES WITH THE FEMORAL NAIL GREATER TROCHANTER (GT)OF THE T2 ALPHA FEMUR ANTEGRADE GT/PF NAILING SYSTEM. SUBJECT: 03-072. (B)(6) 2024: BONE NOT CONSOLIDATED AT 6-MONTH POST-OP CLINIC VISIT. PATIENT IS ON BISPHOSPHONATES AND CHEMOTHERAPY SO SURGEON EXPECTS HEALING TO BE SLOWER THAN NORMAL. SIGNIFICANT CALLUS HAS DEVELOPED AND BONE IN THE PROCESS OF HEALING. PATIENT TO RETURN FOR REPEAT X-RAY IN 2 MONTHS. IF UNION NOT ACHIEVED AT THAT POINT WILL DISCUSS SURGICAL TREATMENT OPTIONS."
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 20010232 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page