Skip to content

T2 ALPHA 2353-4113Recalls & reported problems

Stryker GmbH · Nail Adapter Tibia SPI · Class 2

All models of T2 ALPHA (123)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327358018), received between May 9, 2022 and Feb 16, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation2
  • Residue After Decontamination2
  • Break1
  • Device Difficult to Setup or Prepare1
  • Fracture1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 16, 2024Break; Fracture; Device Difficult to Setup or Prepare; Material Deformation; Break; Fracture; Device Difficult to Setup or Prepare; Material Deformation

    AS REPORTED BY REP: "STRIKE PLATE SNAPPED FROM THE TARGETING ADAPTER WHILE INSERTING THE NAIL (HITTING THE STRIKE PLATE USING THE HAMMER)".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18727661 ↗
  • MalfunctionDec 7, 2023Residue After Decontamination; Residue After Decontamination

    AS REPORTED: "DETECTED METAL CHIPS IN INSERTS FOR T2 ALPHA SPI, AFTER WASHING/CLEANING."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18283060 ↗
  • MalfunctionDec 7, 2023Residue After Decontamination; Material Deformation; Residue After Decontamination; Material Deformation

    AS REPORTED: "DETECTED METAL CHIPS IN INSERTS FOR T2 ALPHA SPI, AFTER WASHING/CLEANING."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18283059 ↗
  • MalfunctionMay 9, 2022Mechanical Jam; Mechanical Jam

    AS REPORTED: "THE SURGEON INSERTED A 9MM T2 ALPHA TIBIA NAIL VIA THE SUPRAPATELLAR APPROACH. SURGEON REAMED TO 10.5, AND MEASURED 320. HE CHOSE TO USE A 300 NAIL. THE CANAL WAS NARROW AND THE SURGEON WOULD HAVE USED AN 8MM NAIL IF AVAILABLE. PROXIMAL SCREWS AND THEN DISTAL SCREWS WERE INSERTED. AFTER DISTAL SCREW INSERTION, SURGEON WAS UNABLE TO DETACH NAIL HOLDING BOLT FROM THE IM ROD. THE SURGEON SPENT 10 MINUTES TRYING TO USE THE BALL TIP SCREW DRIVER TO DISENGAGE THE BOLT. HE WAS UNABLE TO DO SO. VICE GRIPS WERE EVENTUALLY USED TO REMOVE THE SCREW. FORCE AND COUNTER FORCE HAD TO BE USE TO DISENGAGE THE SCREW FROM THE ROD AND REMOVE THE NAIL ADAPTER / NAIL HOLDING BOLT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14340611 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page