Skip to content

NA 2351-0030Recalls & reported problems

Stryker GmbH · Guide Wire Handle · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 07613327361674), received between Jun 25, 2019 and Nov 12, 2024 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component9
  • Fracture6
  • Break5
  • Component Missing3
  • Device Slipped1
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • No Known Impact Or Consequence To Patient4
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 12, 2024Break; Fracture; Break; Fracture

    AS REPORTED: "BROKEN INSTRUMENT." NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20665418 ↗
  • MalfunctionAug 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED: "BOLT ON THE INSTRUMENT IS BROKEN OR HAS COME OFF. BOLT ON THE INSTRUMENT CAN NO LONGER BE USED OR CAN ONLY BE USED TO A LIMITED EXTENT".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20026356 ↗
  • MalfunctionJan 3, 2024Break; Fracture; Detachment of Device or Device Component; Break; Fracture; Detachment of Device or Device Component

    AS REPORTED: "GUIDE WIRE HANDLE BROKE ONE "SCREW" OF HANDLE FELL OUT AND RENDERED HANDLE UNUSABLE. ANOTHER TRAY HAD TO BE OPENED TO COMPLETE THE CASE. THERE WAS A 5 MIN SURGICAL DELAY. WASHED BY HOSPITAL SPD AFTER CASE." ALL DEBRIS WAS REMOVED FROM THE SURGICAL FIELD/PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18438060 ↗
  • MalfunctionNov 13, 2023Break; Fracture; Break; Fracture

    AS REPORTED: "OUR GUIDE WIRE HANDLE IN OUR CONSIGNED IM NAILING BASIC TRAY IS BROKEN. IT IS NOT GRABBING THE WIRE LIKE IT SHOULD."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18123375 ↗
  • MalfunctionAug 30, 2023Component Missing; Component Missing

    AS REPORTED: "DURING SURGERY AN INSTRUMENT (GUIDE WIRE HANDLE) WAS FOUND TO BE UNUSABLE AS IT WAS FOUND TO MISSING A PART. IT WAS NECESSARY TO OPEN AN ADDITIONAL SET TO USE THE HANDLE FROM THAT SET TO FINISH THE SURGERY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17657901 ↗
  • MalfunctionAug 7, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED: "DURING PREPARATION OF THE HANDLE FOR THE GUIDE WIRE (T2 ALPHA SYSTEM), A SMALL BOLT GOT LOST."

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17476018 ↗
  • MalfunctionJul 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED: "DURING A LEFT FEMORAL NAIL PROCEDURE, UPON IMPACTION, 2 SCREWS DISLODGED FROM THE GUIDE WIRE HANDLE. ALL PARTS WERE RETRIEVED, NOTHING LEFT IN THE PATIENT. SURGERY WAS COMPLETED SUCCESSFULLY WITH A DELAY OF APPROXIMATELY 1-2 MINUTES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17375836 ↗
  • MalfunctionApr 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED: "AFTER SURGERY THE SMALL "SCREWS"/ RIVET FELL OUT OF THE T2 ALPHA HANDLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16774964 ↗
  • MalfunctionNov 29, 2022Break; Fracture; Break; Fracture

    AS REPORTED: "THE CUSTOMER WAS USING THE GUIDEWIRE HANDLE, AND HITTING ON IT. SUDDENLY IT BREAKS, AND CAN NOT BE USED ANYMORE." THE SURGEON NEEDED A NEW GUIDE WIRE HANDLE, SO THEY HAD TO FIND A NEW ONE IN ANOTHER KIT FROM ANOTHER SUPPLIER. APPROX. 15 MIN. DELAY. ALL DEBRIS WAS REMOVED. THE HANDLE BROKE IN TWO. CLEAN CUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15883339 ↗
  • MalfunctionDec 8, 2020Fracture; Fracture

    AS REPORTED: "WHEN USING THE GUIDE WIRE HANDLE TO INSERT THE 3.2 X 1000 GUIDE WIRE INTO THE IM CANAL OF THE FEMUR, THE DOCTOR WAS USING THE MALLET TO INSERT THE GUIDE WIRE. WHEN IMPACTING THE GUIDE WIRE THE PIVOT POST THAT HOLDS THE HANDLE TOGETHER BROKE. ALL PIECES WERE RECOVERED AND THERE WERE NO COMPLICATIONS WITH THE CASE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10967836 ↗
  • MalfunctionNov 13, 2020Fracture; Fracture

    WHEN USING THE GUIDE WIRE HANDLE TO INSERT THE 3.2 X 1000 GUIDE WIRE INTO THE IM CANAL OF THE FEMUR, THE DOCTOR WAS USING THE MALLET TO INSERT THE GUIDE WIRE. WHEN IMPACTING THE GUIDE WIRE THE PIVOT POST THAT HOLDS THE HANDLE TOGETHER BROKE. ALL PIECES WERE RECOVERED AND THERE WERE NO COMPLICATIONS WITH THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10838654 ↗
  • MalfunctionSep 30, 2020Component Missing; Component Missing

    GUIDE WIRE HANDLE HAS A MISSING PIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10609068 ↗
  • MalfunctionSep 23, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED: "PRIMARY PROCEDURE, LEFT TIBIAL NAIL. IT WAS REPORTED THAT UPON IMPACTION, A SMALL PIECE (POSSIBLY A NUT) FELL OFF OF THE GUIDE WIRE HANDLE. THE PIECE DID NOT FALL INTO THE PATIENT. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY WITH NO SURGICAL DELAY. REP CONFIRMED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10576721 ↗
  • MalfunctionAug 7, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED: DURING A T2 ALPHA TIBIAL NAIL CASE AT ROYAL PERTH HOSPITAL THE SCRUB NURSE NOTICED ONE OF THE STEEL METAL PLUGS HAD FALLEN OUT FROM THE GUIDE WIRE HANDLE. THE GUIDE WIRE HANDLE WAS NOT NEEDED FOR THE REST OF THE CASE. WITH THE METAL PLUG FALLING OUT IT MADE THE GUIDE WIRE HANDLE UNUSABLE AS IT DIDN'T RETAIN WIRES.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10380014 ↗
  • MalfunctionFeb 7, 2020Component Missing; Detachment of Device or Device Component; Component Missing; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT AS HE WAS USING THE T2 ALPHA BASIC GUIDE WIRE HANDLE HE NOTICED THAT THERE WERE SCREWS MISSING FROM THE HANDLE. THE SURGEON BELIEVES THAT THE DEVICE WAS RETURNED FROM RE-PROCESSING WITH THE SCREWS MISSING. THERE IS NO UNINTENDED METAL IN THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9684865 ↗
  • MalfunctionFeb 5, 2020Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE, THE SURGEON WAS HITTING THE GUIDE WIRE HANDLE AND THE HANDLE BROKE. APPARENTLY THE PIECES FLEW ACROSS THE ROOM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9674103 ↗
  • MalfunctionDec 23, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE STRYKER SALES REPRESENTATIVE REPORTED THAT DURING THE CASE, THE SCREW CAME OUT OF THE T2 ALPHA GUIDE WIRE HANDLE. THE SILVER BOLT CAME OUT INTO THE INCISION. THE SURGEON WAS ABLE TO LOCATE THE ITEM BUT THIS RESULTED IN A SURGICAL DELAY. THE SURGEON WAS NOT USING EXCESSIVE FORCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9511482 ↗
  • MalfunctionJun 25, 2019Disconnection; Disconnection; Device Slipped

    SURGEON WAS USING THE T2 ALPHA GUIDE WIRE GRIPPER, WHILST USING IT A SMALL SCREW FELL OUT. THERE WAS A SLIGHT DELAY OF 5 MINUTES IN THE PROCEDURE AS THE STAFF FETCHED ANOTHER GUIDE WIRE HANDLE. STRYKER REP REPORTED THAT THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8732805 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →