NA 2361-5050SRecalls & reported problems
Stryker GmbH · Advanced Locking Screw · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327381559), received between Apr 8, 2020 and Apr 2, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Device Dislodged or Dislocated1
- Fracture1
- Mechanical Jam1
- Migration1
Patient problems
- Failure of Implant1
- Foreign Body In Patient1
- Injury1
- No Clinical Signs, Symptoms or Conditions1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "ORIF WAS PERFORMED FOR FEMORAL CONDYLE FRACTURE ON (B)(6) 2024. HOWEVER, IT WAS FOUND THAT FIXATION WAS NOT ACHIEVED ON (B)(6) 2024. THUS, REVISION SURGERY TO REMOVE THE DEVICES WAS PERFORMED ON (B)(6) 2024. THE DOCTOR TOLD THAT THE ORIGINAL FRACTURE WAS SEVERE, AND ALTHOUGH THE DEVICES WERE IMPLANTED IN HOPES OF REDUCING AND MAINTAINING THE FRACTURE, THE FRACTURE PART REMAINED UNSTABLE AFTER SURGERY."
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 19026635 ↗MalfunctionNov 15, 2022Break; Fracture; Mechanical Jam; Break; Fracture; Mechanical Jam
AS REPORTED: "DURING A T2 ALPHA GT PROCEDURE, AN ADVANCED LOCKING SCREW WAS USED, BUT IT BECAME STUCK WITH THE NAIL AND COULD NOT BE PULLED OUT OR ADVANCED. IT GOT STUCK IN THE MIDDLE OF INSERTION AND COULD NOT BE PULLED OUT DESPITE VARIOUS ATTEMPTS TO REMOVE IT. THE PROTRUDING PART WAS CRUSHED WITH A CARBIDE DRILL AS A LAST RESORT. THE REMAINDER OF THE SCREW WAS LEFT IN THE BODY." ADDITIONALLY, IT WAS REPORTED THAT THERE WAS A 60 MINUTE SURGICAL DELAY.
Patient problems: Foreign Body In Patient
MAUDE report 15797423 ↗MalfunctionFeb 18, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED: "THE INITIAL OPERATION FOR TIBIAL TRUNK SURGERY WAS PERFORMED ON (B)(6) 2020. SINCE T2 ALPHA ADVANCED LOCKING SCREW WAS COMING OFF (SINCE UNKNOWN DATE). THE PATIENT WAS CAUSING THE INFECTION, AND THERE MAY BE SOME CAUSAL RELATIONSHIP."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11346641 ↗InjuryApr 8, 2020Migration; Migration
THIS REPORT IS FOR THE SECOND SCREW BACK-OUT ("THE OTHER 2 DISTAL SCREWS"). AS REPORTED: "THE MOST DISTAL T2 ALPHA SCREW IN THE TIBIA NAIL BACKED OUT AND HAD TO BE REMOVED. THE SCREW WAS AN ADVANCED LOCKING SCREW AND WAS REMOVED IN CLINIC. TWO DAYS LATER, THE OTHER 2 DISTAL SCREWS (ADVANCED LOCKING SCREWS) ALSO BACKED OUT. THIS EVENT HAPPENED 2 AND 2.5 WEEKS POST OP".
Patient problems: Failure of Implant; Injury
MAUDE report 9941480 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page