The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327381634), received between Nov 23, 2021 and Apr 18, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Difficult to Setup or Prepare1
  • Device Dislodged or Dislocated1
  • Failure to Advance1
  • Fracture1
  • Material Deformation1
  • Mechanical Jam1
  • Migration1

Patient problems

  • Failure of Implant3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 18, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED BY THE SALES REP THAT A RETROGRADE T2 ALPHA FEMUR NAIL AND AN UNKNOWN SCREW BROKE POSTOPERATIVELY. REVISION SURGERY TOOK PLACE ON (B)(6) 2024.

    Patient problems: Failure of Implant

    MAUDE report 19134983 ↗
  • MalfunctionSep 1, 2022Device Difficult to Setup or Prepare; Failure to Advance; Material Deformation; Device Difficult to Setup or Prepare; Failure to Advance; Material Deformation

    AS REPORTED: ADVANCED LOCKING SCREW CROSS THREADED INTO OBLIQUE HOLE. SCREW WOULD NOT ADVANCE (OR BACK OUT) WITH POWER OR HAND TO THE POINT OF STRIPPING THE SCREW. EVENTUALLY WAS ABLE TO PUT ON SHORT CAPTURE SCREWDRIVER TO BARELY REMOVE SCREW. REPLACED WITH A STANDARD LOCKING SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15335673 ↗
  • MalfunctionMar 31, 2022Mechanical Jam; Mechanical Jam

    T2 ALPHA ADVANCED LOCKING SCREW WAS COLD WELDED TO THE NAIL. EVENTUALLY IT WAS ABLE TO BE REMOVED, AND WAS REPLACED WITH A STANDARD LOCKING SCREW. ADDITIONALLY, IT WAS REPORTED; THE PROCEDURE WAS DELAYED BUT THE LENGTH OF THE DELAY IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13964480 ↗
  • MalfunctionMar 8, 2022Migration; Migration

    IT WAS REPORTED: "ONE MONTH POSTOPERATIVE EXAMINATION SHOWED BACKOUT OF THE ADVANCED LOCKING SCREW. AT THE MOMENT, NO ADDITIONAL INFORMATION IS AVAILABLE AND NO REVISION IS PLANNED."

    Patient problems: Failure of Implant

    MAUDE report 13693463 ↗
  • InjuryNov 23, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED: "THE DOCTOR IMPLANTED A T2 ALPHA TIBIA NAIL. AFTER A FEW MONTHS THE ADVANCED LOCKING SCREWS BACKED OUT." AS A RESULT THE IMPLANTS WERE EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12860045 ↗