Synchro2 S2SPP14215Recalls & reported problems

Stryker Corporation · SUPPORT PRE-SHAPED Guidewire with Hydrophilic Coating · Class 2

All models of Synchro2 (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
8
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07613327459654), received between Feb 4, 2020 and Apr 26, 2024 — classified as 8 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (12 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 18, 2024Open, Classified

    Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

    Z-1941-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture12
  • Peeled/Delaminated3

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Foreign Body In Patient4
  • Embolism/Embolus1
  • No Known Impact Or Consequence To Patient1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 26, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE, THE TIP OF THE SUBJECT GUIDEWIRE WAS BROKEN AND REMAINED IN THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 19197260 ↗
  • MalfunctionSep 15, 2023Fracture; Peeled/Delaminated; Fracture; Peeled/Delaminated

    THE SUBJECT DEVICE WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED THAT THE SUBJECT GUIDEWIRE HAD PTFE (POLYTETRAFLUOROETHYLENE) COATING PEELING ISSUE AND THE SUBJECT GUIDEWIRE DISTAL TIP WAS BROKEN/FRACTURED DURING USE. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17761590 ↗
  • InjuryAug 10, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A NEUROVASCULAR PROCEDURE THE DISTAL PART OF THE SUBJECT GUIDEWIRE BROKE DURING USE. STENT THROMBECTOMY WAS PERFORMED TO REMOVE THE BROKEN GUIDEWIRE FROM THE PATIENTS BODY. THE BROKEN PIECE WAS REMOVED SUCCESSFULLY WHICH DID NOT INFLUENCE THE OUTCOME OF THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17502863 ↗
  • InjuryFeb 16, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE (SUBJECT DEVICE) WAS SHEARED OFF WHEN CROSSING OCCLUSION DURING MANIPULATION OF THE WIRE ATTEMPTING TO CROSS THROMBUS, PHYSICIAN NOTICED THAT HE HAD LOST 1 TO 1 TORQUE RESPONSE. PART OF THE GUIDEWIRE (SUBJECT DEVICE) WAS LEFT IN THE PATIENT AND STENT PLACEMENT WAS PERFORMED AS A MEDICAL INTERVENTION. PATIENT STATUS IS STABLE. NIHSS (NATIONAL INSTITUTES OF HEALTH STROKE SCALE) PRE PROCEDURE WAS 23 AND POST PROCEDURE WAS 7. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Foreign Body In Patient

    MAUDE report 16391604 ↗
  • InjuryNov 28, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE SUBJECT GUIDEWIRE TORN OFF DURING USE. THE DEVICE WAS REMOVED WITH THE HELP OF A SNARE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO CLINICAL CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15869156 ↗
  • MalfunctionDec 20, 2021Fracture; Peeled/Delaminated; Fracture; Peeled/Delaminated

    IT WAS REPORTED THAT DURING PREPARATION, THE GUIDEWIRE (SUBJECT DEVICE) HAD A "PIG-TAIL" SHAPED TIP. WHEN THE GUIDEWIRE (SUBJECT DEVICE) WAS USED AGAIN, THE INSERTION AID WAS USED. FOR THIS PURPOSE, IT WAS PUSHED OVER THE WIRE FROM BEHIND. IT WAS NOTICED THAT PARTICLES OF THE OUTER COATING OF THE GUIDE WIRE (SUBJECT DEVICE) WERE COMING LOOSE. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE WITHOUT ANY CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Thrombosis/Thrombus

    MAUDE report 13049575 ↗
  • InjuryDec 13, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO HOSPITAL WITH THE MAJOR STROKE. THE SUBJECT WIRE ALONG WITH OTHER DEVICES WERE USED TO REMOVE THE HARD CLOT LOCATED AT THE LEFT POSTERIOR VIA THE INTERNAL CAROTID ARTERY. WHILE PROBING, THE TIP OF THE WIRE (SUBJECT DEVICE) GOT STUCK IN THE CLOT/STENOSIS AND WHILE TWISTING ON PROXIMAL SIDE THE SUBJECT WIRE WAS TORN OFF AND CAUSED A 120-MINUTE SURGICAL DELAY. THE ASS MEDICAL INTERVENTION WAS PERFORMED TO RETRACT THE BROKEN SUBJECT WIRE BUT UNSUCCESSFUL AS IT REMAINED INSIDE THE PATIENT. IN THE PROCESS, A PROBABLY HARD THROMBUS WAS PASSED. IT WAS LATER REPORTED THAT THE PATIENT HAD PASSED AWAY DUE TO PRESENTING ISCHEMIC STROKE CONDITION PRIOR TO TREATMENT AND THE OUTCOME OF PATIENT DEATH WAS NOT RELATED TO THE SUBJECT WIRE DEVICE. UPDATE : IT WAS CONFIRMED THAT THE PROBABLY HARD THROMBUS WAS PASSED THAT OCCURRED IN THE SAME TERRITORY AS THE EMBOLISM. THE PHYSICIAN TRIED TREAT THE THROMBUS BY USING AN ASPIRATION CATHETER FIRST. AFTER THAT DIDN¿T WORK THEY TRIED USING A STENT RETRIEVER. ACCORDING TO THE PHYSICIAN INDICATED THAT THE HARD THROMBUS WAS NOT RELATED TO THE SUBJECT SYNCHRO.

    Patient problems: Embolism/Embolus; Foreign Body In Patient

    MAUDE report 12984972 ↗
  • InjurySep 27, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING ONYX EMBOLIZATION PROCEDURE OF A ARTERIOVENOUS FISTULA, THE DISTAL END OF THE SUBJECT GUIDEWIRE BROKE OFF INSIDE THE PATIENT. THE SUBJECT GUIDEWIRE TIP REMAINS INSIDE THE PATIENT AND IT WAS EMBOLIZED INTO THE FISTULA. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO SNARE DEVICE WAS USED AND THE PATIENT WAS DOING WELL. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 12533143 ↗
  • MalfunctionJan 20, 2021Fracture; Fracture

    INITIALLY IT WAS REPORTED THAT THE GUIDEWIRE DISTAL TIP WAS NOTED TO BE STRETCHED WHEN REMOVED FROM THE PATIENT ANATOMY. THE GUIDEWIRE (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED THAT THE DISTAL TIP OF THE SUBJECT GUIDEWIRE WAS BROKEN/FRACTURED DURING USE. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11206120 ↗
  • InjuryDec 22, 2020Fracture; Peeled/Delaminated; Fracture; Peeled/Delaminated

    IT WAS REPORTED THAT DURING THE PROCEDURE THE SUBJECT GUIDE WIRE BROKE INSIDE THE PATIENT. A SNARE DEVICE WAS USED TO RETRACT THE BROKEN PIECE OF THE SUBJECT GUIDE WIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11059831 ↗
  • MalfunctionDec 9, 2020Fracture; Fracture

    THE GUIDEWIRE (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED THAT THE DISTAL TIP OF THE SUBJECT GUIDEWIRE WAS BROKEN/FRACTURED DURING USE. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10974153 ↗
  • InjuryFeb 4, 2020Fracture; Fracture

    IT WAS REPORTED THAT DURING PROCEDURE TO TREAT A TO TREAT BASILAR STENOSIS, THE PHYSICIAN WAS GOING TO IMPLANT A STENT OVER A GUIDEWIRE DISTAL PART (SUBJECT DEVICE) AND THE GUIDEWIRE BROKE OFF IN THE BASILAR ARTERY IN THE PATIENT. THE PHYSICIAN SUCCESSFULLY SNARED THE DISTAL PIECE OF GUIDEWIRE THAT BROKE OFF. THE STENTING PROCEDURE WAS ABORTED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9668879 ↗

Other Wire, guide, catheter devices

All Wire, guide, catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page