The Synchro2 Guidewires are a steerable guidewire family… recallZ-1941-2024 · Stryker Neurovascular · initiated Apr 18, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

FDA root cause category: Component design/selection

What the company is doing

On 4/29/24, recall notices were mailed to Supply Chain Management/ Recall Coordinator/ Inventory Manager customers who were asked to do the following: 1) When using impacted guidewires, please do not backload the guidewire, follow the Instructions for Use (IFU) provided with the device, and do not use the guidewire if you see any visible signs of coating damage. 2) Circulate this customer notification internally to all interested/affected parties. 3) Inform firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that the firm can inform the recipients appropriately. 4) Inform firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: [email protected]) 5) Complete and return the customer response form via email to: [email protected]. 6) Contact the firm if you would like a replacement of the guidewire with a new optional introducer accessory.

Recall details

Recall number
Z-1941-2024
Status
Open, Classified
Initiated by the company
Apr 18, 2024
Posted by the FDA
May 31, 2024
Recalling firm
Stryker Neurovascular
Quantity
373,246
Product code
DQX
FDA submissions
K053268
Distribution
Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, JORDAN, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UK.

Product description

The Synchro2 Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014 in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque.

Code information (models, lots, serial numbers)

Show the full code information (12k characters)

Catalog/UDI-DI/Lots: M00326110/04546540688736/0000064108, 0000065017, 0000065616, 0000066470, 0000066581, 0000067561, 0000067801, 0000068246, 0000069276, 0000071094, 0000071841, 0000073548, 0000075207, 0000077751, 0000078371, 0000078837, 0000080548, 0000080988, 0000082965, 0000083195, 0000086130, 0000088254, 0000092394, 0000092395, 0000092396, 0000095029, 0000097492, 0000098543, 0000104203, 0000111901, 0000122898, 0000123072, 0000123529, 0000126054, 0000126497, 0000126852, 0000134099, 0000135605, 0000135804, 0000144373, 0000145738, 0000145739, 0000146146, 0000146147, 0000150989, 0000153606, 0000162889, 0000162890, 0000162891, 0000172273, 0000174770, 0000175634, 0000175636, 0000175637; M00326320/04546540688750/0000105764, 0000109900, 0000181399; M00326010/04546540688729/0000063878, 0000063985, 0000064269, 0000064281, 0000064309, 0000064357, 0000064360, 0000064370, 0000064796, 0000065004, 0000065161, 0000065227, 0000065269, 0000065449, 0000065450, 0000065733, 0000065760, 0000065781, 0000065819, 0000065891, 0000066007, 0000066076, 0000066129, 0000066311, 0000066337, 0000066383, 0000066490, 0000066583, 0000066931, 0000066988, 0000067168, 0000067331, 0000067453, 0000067466, 0000067566, 0000067655, 0000067710, 0000067762, 0000067898, 0000068187, 0000068245, 0000068428, 0000068556, 0000069385, 0000069506, 0000069549, 0000069573, 0000069755, 0000069788, 0000069918, 0000070038, 0000070039, 0000070422, 0000070863, 0000071213, 0000071374, 0000071375, 0000071839, 0000071840, 0000073408, 0000073409, 0000073514, 0000073888, 0000074665, 0000074666, 0000074951, 0000074952, 0000075877, 0000076063, 0000076064, 0000076161, 0000076162, 0000076379, 0000076454, 0000077608, 0000077609, 0000077610, 0000078110, 0000078111, 0000078271, 0000078369, 0000078877, 0000078902, 0000080915, 0000080987, 0000081397, 0000081454, 0000081670, 0000082734, 0000082735, 0000084656, 0000084657, 0000084861, 0000084864, 0000085365, 0000085953, 0000086129, 0000086175, 0000086200, 0000086337, 0000087870, 0000087871, 0000087872, 0000088367, 0000088651, 0000088652, 0000090311, 0000091480, 0000091715, 0000092004, 0000092839, 0000093332, 0000093534, 0000094776, 0000094818, 0000095080, 0000095172, 0000096911, 0000096912, 0000097447, 0000097951, 0000098059, 0000099701, 0000099702, 0000099953, 0000100863, 0000100925, 0000100986, 0000102835, 0000102934, 0000103038, 0000103039, 0000104204, 0000104522, 0000104523, 0000106631, 0000106632, 0000106718, 0000107496, 0000107498, 0000108264, 0000108487, 0000109375, 0000109781, 0000109782, 0000109899, 0000111454, 0000111900, 0000112880, 0000112881, 0000113135, 0000113476, 0000113477, 0000113681, 0000114843, 0000114844, 0000114845, 0000115168, 0000116702, 0000116703, 0000117042, 0000117043, 0000121693, 0000121695, 0000122112, 0000122257, 0000122470, 0000123530, 0000123902, 0000123982, 0000123984, 0000125041, 0000125147, 0000125714, 0000125715, 0000126047, 0000126048, 0000126049, 0000126496, 0000128252, 0000128618, 0000128619, 0000128620, 0000129206, 0000129207, 0000130642, 0000130824, 0000130864, 0000130865, 0000132955, 0000133180, 0000133181, 0000134097, 0000134098, 0000137241, 0000137242, 0000138582, 0000138583, 0000139438, 0000139439, 0000139440, 0000139441, 0000140683, 0000140685, 0000140686, 0000140687, 0000146927, 0000146928, 0000147307, 0000150096, 0000150097, 0000150879, 0000150880, 0000150988, 0000152801, 0000152802, 0000152978, 0000153316, 0000153369, 0000157899, 0000158129, 0000158130, 0000158341, 0000158800, 0000159841, 0000159842, 0000161219, 0000161220, 0000162107, 0000162894, 0000166350, 0000166351, 0000166352, 0000166353, 0000170244, 0000172271, 0000172272, 0000172302, 0000172304, 0000172599, 0000172600, 0000172620, 0000172621, 0000173240, 0000173242, 0000174284, 0000174287, 0000174768, 0000174769, 0000175639, 0000175710, 0000175750, 0000176474, 0000176475, 0000176833, 0000176834, 0000177999, 0000178005, 0000180681; M00326310/04546540688743/0000065570, 0000067169, 0000068247, 0000079906, 0000090099, 0000094378, 0000100182.0000108962, 0000116675, 0000118886, 0000128525, 0000135805, 0000147879, 0000147987, 0000157735, 0000157736; M00326420/04546540688774/0000063779, 0000064566, 0000065241, 0000065646, 0000065762, 0000066263, 0000066472, 0000066582, 0000066989, 0000068096, 0000070222, 0000070864, 0000072031, 0000073134, 0000076285, 0000077556, 0000079694, 0000080326, 0000080455, 0000081868, 0000083175, 0000083952, 0000084099, 0000085613, 0000089328, 0000090670, 0000091905, 0000093818, 0000094819, 0000096587, 0000097110, 0000098541, 0000101652, 0000103036, 0000108815, 0000108961, 0000109973, 0000111204, 0000114401, 0000116311, 0000118674, 0000121398, 0000122490, 0000124885, 0000125790,0000135316, 0000135604, 0000138236, 0000138237, 0000141438, 0000147437, 0000147587, 0000148962, 0000148964, 0000151850, 0000152226, 0000155067, 0000155661, 0000161018, 0000163156, 0000172923, 0000172924, 0000174675, 0000176017, 0000176114, 0000176473, 0000176525; M00326520/04546540688798/0000066684, 0000073892, 0000090098, 0000105762, 0000122738, 0000151278; M00326410/04546540688767/0000063497, 0000063501, 0000063520, 0000063568, 0000063716, 0000063767, 0000063879, 0000063894, 0000063905, 0000063933, 0000064072, 0000064113, 0000064410, 0000064420, 0000064482, 0000064486, 0000064487, 0000064527, 0000064565, 0000064589, 0000064590, 0000064735, 0000064736, 0000064797, 0000064882, 0000064927,0000064960,0000064961, 0000065005, 0000065107, 0000065118, 0000065281, 0000065370, 0000065515, 0000065571, 0000065583, 0000065643, 0000065761, 0000065820, 0000065837, 0000065900, 0000066008, 0000066075, 0000066130, 0000066262, 0000066384, 0000066385, 0000066471,0000066552, 0000066644, 0000066824, 0000066962, 0000067082, 0000067170, 0000067332, 0000067456, 0000067567, 0000067653, 0000067654, 0000067711, 0000067712, 0000067829, 0000068248, 0000068627, 0000069225, 0000069730, 0000069790, 0000069919, 0000070220, 0000070221, 0000070423, 0000070424, 0000070640, 0000070641, 0000071095, 0000071096, 0000071378, 0000072028, 0000072029, 0000072244, 0000072277, 0000073133, 0000073549, 0000073550, 0000074431, 0000074451, 0000075208, 0000075209, 0000075553, 0000075554, 0000075915, 0000076455, 0000077386, 0000077387, 0000077389, 0000077554, 0000077555, 0000078272, 0000078273, 0000078372, 0000079396, 0000080021, 0000080914, 0000081395, 0000081697, 0000081741, 0000081742, 0000082781, 0000082985, 0000082986, 0000083855, 0000083856, 0000085274, 0000085367, 0000085368, 0000085369, 0000085635, 0000085636, 0000085724, 0000085952, 0000086425, 0000086538, 0000086539, 0000086540, 0000086939, 0000086940, 0000087549, 0000087550, 0000087582, 0000087635, 0000088255, 0000088811, 0000088890, 0000089000, 0000089068, 0000089069, 0000089523, 0000090313, 0000090391, 0000090548, 0000090549, 0000091107, 0000091372, 0000091477, 0000091478, 0000091479, 0000092564, 0000092594, 0000092702, 0000092838, 0000093535, 0000093536, 0000094533, 0000094737, 0000094742, 0000094931, 0000094935, 0000095077, 0000095424, 0000095426, 0000095429, 0000095475, 0000095836, 0000095874, 0000096624, 0000096629, 0000096630, 0000096631, 0000096772, 0000096914, 0000097107, 0000097108, 0000097505, 0000097695, 0000097696, 0000097784, 0000097841, 0000097842, 0000099524, 0000099527, 0000099660, 0000100303, 0000100602, 0000100988, 0000101377, 0000101378, 0000101621, 0000101622, 0000101651, 0000102589, 0000102793, 0000102794, 0000102795, 0000102796, 0000103037, 0000103809, 0000104202, 0000104524, 0000104574, 0000104973, 0000104974, 0000105253, 0000105351, 0000105763, 0000105956, 0000105957, 0000106028, 0000106405, 0000106411, 0000106412, 0000107495, 0000107934, 0000107935, 0000108145, 0000108433, 0000108813, 0000109901, 0000110233, 0000110879, 0000110880, 0000111101 ,0000111453, 0000112240, 0000112285, 0000112709, 0000112822, 0000112823, 0000112844, 0000113136, 0000113690, 0000113790, 0000113791, 0000114166, 0000114167, 0000114170, 0000115255, 0000115598, 0000115599, 0000115668 ,0000115669, 0000115670, 0000115926, 0000115927, 0000115965, 0000115966, 0000117044, 0000117679, 0000117830, 0000117831, 0000118137, 0000118138, 0000118202, 0000118415, 0000118416, 0000118675, 0000118804, 0000119600, 0000119601, 0000120211, 0000121051, 0000121106, 0000121215, 0000121216, 0000123073, 0000123148, 0000123149, 0000123406, 0000123493, 0000123526, 0000123527, 0000123983, 0000124070, 0000124158, 0000124159, 0000124212, 0000124213, 0000124629, 0000124630, 0000124637, 0000124638, 0000124846, 0000124847, 0000124848, 0000125413, 0000125417, 0000125696, 0000125717, 0000125718, 0000125788, 0000125789, 0000126853, 0000126923, 0000126924, 0000127370, 0000127371, 0000127750, 0000127751, 0000128386, 0000128387, 0000128524, 0000129626, 0000129628, 0000129839, 0000129840, 0000129900, 0000129901, 0000131302, 0000131303, 0000131335, 0000131336, 0000131720, 0000131721, 0000131722, 0000132105, 0000132106, 0000132107, 0000132150, 0000132151, 0000132152, 0000132739, 0000133393, 0000133394, 0000133846, 0000133847, 0000133931, 0000133932, 0000134092, 0000134093, 0000136250, 0000136251, 0000136411, 0000136412, 0000136676, 0000136677, 0000136918, 0000136919, 0000136922, 0000137320, 0000137321, 0000138750, 0000140167, 0000140175, 0000140308, 0000140564, 0000140568, 0000140977, 0000141439, 0000141440, 0000141441, 0000144374, 0000144831, 0000144832, 0000145094, 0000147187, 0000147188, 0000147383, 0000147438, 0000147439, 0000147588, 0000147983, 0000147984, 0000148347, 0000148348, 0000148349, 0000148350, 0000148351, 0000148889, 0000148892, 0000148963, 0000149417, 0000149418, 0000150098, 0000151282, 0000151283, 0000151284, 0000151382, 0000151383, 0000151679, 0000151849, 0000152229, 0000152374, 0000152519, 0000152520, 0000152521, 0000152800, 0000153607, 0000153694, 0000153695, 0000153972, 0000153984, 0000154307, 0000154308, 0000155015, 0000155066, 0000155069, 0000155663, 0000155664, 0000156793, 0000157431, 0000157433, 0000157434, 0000157890, 0000157891, 0000158805, 0000158806, 0000159540, 0000159993, 0000159995, 0000160373, 0000160380, 0000160381, 0000160432, 0000161017, 0000163717, 0000164441, 0000164660, 0000164661, 0000165475, 0000169345, 0000169346, 0000170245, 0000171218, 0000171219, 0000172527, 0000172529, 0000172530, 0000172863, 0000172864, 0000173164, 0000173935, 0000173936, 0000173937, 0000174288, 0000174291, 0000174352, 0000174353, 0000174372, 0000174404, 0000174405, 0000174406, 0000174407, 0000174408, 0000176115, 0000176117, 0000176118, 0000176329, 0000176330, 0000176827, 0000176828, 0000176829, 0000177034, 0000177035, 0000177078, 0000177079, 0000177080, 0000177521, 0000177527, 0000177533, 0000180797, 0000180798; M00326510/04546540688781/0000063768, 0000063886, 0000064191, 0000064640, 0000065066, 0000066131, 0000066729, 0000066911, 0000067656, 0000067658, 0000067683, 0000068097, 0000069226, 0000069386, 0000069867, 0000070229, 0000070642, 0000071089, 0000073894, 0000074671, 0000075916, 0000077169, 0000077170, 0000078109, 0000079633, 0000081204, 0000081875, 0000081920, 0000082489, 0000083953, 0000084100, 0000086865, 0000087972, 0000089737, 0000090550, 0000091906, 0000093819, 0000094029, 0000094030, 0000097109, 0000097917, 0000099212, 0000101833, 0000102061, 0000103708, 0000105594, 0000107290, 0000109972, 0000111203, 0000111220, 0000114545, 0000116312, 0000116339, 0000118803, 0000118885, 0000121401, 0000121582, 0000121918, 0000124886, 0000126024, 0000126027, 0000134094, 0000135305, 0000139099, 0000139100, 0000142534, 0000146227, 0000146388, 0000146389, 0000146508, 0000146509, 0000150094, 0000150870, 0000152518, 0000152979, 0000153296, 0000153309, 0000156613, 0000157730, 0000158801, 0000160739, 0000163715, 0000164565, 0000170923, 0000170924, 0000170925, 0000173347, 0000173348, 0000173676, 0000173677, 0000174158, 0000174374, 0000175063, 0000177059, 0000177060, 0000177066; S2SPP14215/07613327459654/0000064643, 0000065728, 0000066913, 0000067173, 0000078582, 0000084347, 0000092593, 0000097916, 0000108963, 0000121105, 0000139102, 0000141827, 0000144378, 0000146510, 0000164584; S2SPP14300/07613327459661/0000084348; S2SPS14215/07613327459630/0000064656, 0000065327, 0000065644, 0000067080, 0000067171, 0000067663, 0000068249, 0000072389, 0000084516, 0000084517, 0000091104, 0000095875, 0000100603, 0000103807, 0000109593, 0000113421, 0000114789, 0000118119, 0000121583, 0000122737, 0000125039, 0000126355, 0000128621, 0000139101, 0000141394, 0000143124, 0000143346, 0000147308, 0000176835; S2SPS14300/07613327459647/0000065435, 0000084518, 0000094339, 0000118540, 0000128617, 0000140967, 0000146841

Devices named in this recall

12 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 94543).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page