GAMMA 8425-0320SRecalls & reported problems

Stryker GmbH · Long Nail, RIGHT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327479515), received between Mar 20, 2025 and Mar 23, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 28, 2025Class IIOpen, Classified

    Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.

    Z-1914-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Malposition of Device1
  • Positioning Failure1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Perforation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 23, 2026Fracture

    AS REPORTED: "THE DISTRIBUTOR CONTACTED THE SALES REPRESENTATIVE TO REPORT THAT A GAMMA4 NAIL USED ON A PATIENT HAD BROKEN."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24670110 ↗
  • InjuryDec 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT "DURING INSERTION OF A GAMMA 4 LONG NAIL FOR A SUBTROCHANTERIC FEMORAL FRACTURE, THE NAIL TIP PERFORATED THE ANTERIOR ASPECT OF THE FEMUR AT THE UPPER FEMORAL CONDYLE. THE ENTRY WAS REPOSITIONED ANTERIORLY AND THE NAIL WAS REINSERTED DEEPLY. THE SURGERY WAS COMPLETED. THE DEVICE IS IMPLANTED IN THE PATIENT'S BODY".

    Patient problems: Insufficient Information

    MAUDE report 23783832 ↗
  • MalfunctionMar 20, 2025Positioning Failure; Malposition of Device; Positioning Failure; Malposition of Device

    AS REPORTED: "OPENED THE RIGHT NAIL OF GAMMA 4 DIAMETER 10MM / LENGTH 320MM. LUG SCREW INTERFERENCE CHECK WAS PERFORMED. AFTER INSERTING THE NAIL AND LAG SCREW, THE DISTAL DEVICE WAS ATTACHED, AND AFTER ADJUSTING THE HEIGHT, THE MOST DISTAL CIRCULAR HOLE WAS DRILLED, BUT AFTER PASSING THROUGH THE FRONT HOLE, THE DRILL DID NOT PASS THROUGH DUE TO INTERFERENCE INSIDE THE NAIL. AFTER THAT, WE TRIED THE DISTAL MOST ROUND HOLE AND THE ELLIPTICAL HOLE, BUT THE DRILLS DID NOT PASS THROUGH ALL OF THEM. WHEN THE C-ARM WAS INSERTED FROM RIGHT BESIDE THE DISTAL END OF THE NAIL TO SWITCH TO RADIOLUCENT, IT WAS DISCOVERED THAT THE CUT SURFACE AT THE DISTAL END OF THE NAIL WAS NOT FORWARD, BUT BACKWARD. AFTER REMOVING THE NAIL, A NAIL ONE SIZE SHORTER (300 MM) WAS REINSERTED. SURGERY WAS COMPLETED. AFTER SURGERY, THE DISTAL DEVICE AND NAIL WERE INSTALLED AND CALIBRATED, AND THE DRILL DID NOT PASS".

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Perforation

    MAUDE report 21659994 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →