Why it was recalled

Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.

FDA root cause category: Process design

What the company is doing

Stryker notified consignees on about 05/28/2025 via FedEx letter. Consignees were instructed to review inventory and immediately quarantine any affected units on hand, and complete and return the Business Reply Form documenting receipt and product segregation. Upon receipt of the response form, Stryker will contact consignees to arrange for return of the affected units. If affected products have been further distributed, consignees were instructed to notify those customers and let Stryker know.

Recall details

Recall number
Z-1914-2025
Status
Open, Classified
Initiated by the company
May 28, 2025
Posted by the FDA
Jun 6, 2025
Recalling firm
Stryker GmbH
Product code
HSB
FDA submissions
K213328
Distribution
Worldwide - US Nationwide distribution in the states of AL, FL, NJ, OR and the countries of Australia and Japan.

Product description

Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.

Code information (models, lots, serial numbers)

Catalog Number: 84250320S. GTIN: 07613327479515. Lot Number: K1BC9B7.