iBur 8431-013-040DCRecalls & reported problems

Stryker Corporation · 4.0mm Coarse Diamond Round, Distal Bend · Class 2

All models of iBur (15)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07613327501155), received between Apr 15, 2022 and Oct 23, 2025 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Fracture” (8 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 24, 2024Class IICompleted

    Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

    Z-2728-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture8
  • Overheating of Device3

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2025Fracture; Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) - (B)(6) 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368685 ↗
  • MalfunctionJul 29, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FRACTURE. 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22648125 ↗
  • MalfunctionJan 22, 2024Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18557311 ↗
  • MalfunctionJun 23, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR BROKE. IT WAS ALSO REPORTED THE EVENT CAUSED A DELAY OF 5 MINUTES. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17195563 ↗
  • MalfunctionMay 26, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED, THAT DURING A SURGICAL PROCEDURE. THE BUR OVERHEATED. IT WAS ALSO REPORTED, THERE WERE NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED, THAT THE EVENT CAUSED A DELAY OF A FEW MINUTES. IT WAS ALSO REPORTED, THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17011160 ↗
  • MalfunctionSep 16, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE INSTANCE OF THE BURS BREAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15441871 ↗
  • MalfunctionJul 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THREE BURS BROKE DURING THE SAME PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE BURS WHICH WAS REPORTED TO HAVE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14982006 ↗
  • MalfunctionJul 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THREE BURS BROKE DURING THE SAME PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE BURS WHICH WAS REPORTED TO HAVE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14981944 ↗
  • MalfunctionJul 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THREE BURS BROKE DURING THE SAME PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE BURS WHICH WAS REPORTED TO HAVE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14981881 ↗
  • MalfunctionApr 15, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, AS A RESULT OF LACK OF IRRIGATION, TWO BURS OVERHEATED. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE INSTANCE OF A BUR OVERHEATING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14117839 ↗
  • MalfunctionApr 15, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, AS A RESULT OF LACK OF IRRIGATION, TWO BURS OVERHEATED. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE INSTANCE OF A BUR OVERHEATING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14117750 ↗

FDA clearance and classification

The FDA regulates product code HBE (drills, burrs, trephines & accessories (simple, powered)) under 21 CFR 882.4310 as a Class II device. 21 CFR 882.4310 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the iBur 8431-013-040DC

Has the iBur 8431-013-040DC been recalled?

Yes. 1 FDA recall record names this device's identifier, and the company has completed it but the FDA has not yet closed it. The recall, Z-2728-2024 (Class II), began on Jul 24, 2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the iBur 8431-013-040DC?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Apr 15, 2022 and Oct 23, 2025. In those reports, the most common specific device problems are “Fracture” (8) and “Overheating of Device” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the iBur 8431-013-040DC?

None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the iBur 8431-013-040DC, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand iBur, model 8431-013-040DC). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.