iBur 8431-013-050DCRecalls & reported problems
Stryker Corporation · 5.0mm Coarse Diamond Round, Distal Bend · Class 2
All models of iBur (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 07613327501186), received between Sep 14, 2022 and Jan 23, 2026 — classified as 1 injury, 18 malfunctions.
The most frequently reported specific device problem is “Fracture” (18 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 24, 2024Class IICompleted
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Z-2729-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture18
- Overheating of Device1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Fracture
MalfunctionAug 28, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE BUR BROKE AT THE CUTTING END. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20091022 ↗MalfunctionMay 22, 2024Fracture; Fracture
PER MEDWATCH MW5153659; IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE BUR BROKE AND ALL BUR FRAGMENTS WERE RETRIEVED SUCCESSFULLY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19375885 ↗MalfunctionFeb 20, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, ONE OF THE BURS BROKE. IT WAS ALSO REPORTED THAT THE EVENT CONTRIBUTED TO A SURGICAL DELAY OF 10 MINUTES. IT WAS FURTHER REPORTED THAT NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION OCCURRED DURING THE PROCEDURE. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18738148 ↗MalfunctionJan 22, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18556911 ↗InjuryOct 9, 2023Fracture; Fracture
PER MEDWATCH REPORT#: MW5145184; IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR BROKE. IT WAS ALSO REPORTED THE EVENT CAUSED AN UNKNOWN SERIOUS INJURY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17894147 ↗MalfunctionSep 12, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A ANTERIOR CERVICAL DISCECTOMY AND FUSION(ACDF) SURGICAL PROCEDURE, THE BUR BROKE WHILE DRILLING. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE EVENT CAUSED A 2 MINUTE DELAY TO THE PROCEDURE. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17731408 ↗MalfunctionJul 12, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE BUR BREAKS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17306814 ↗MalfunctionJul 12, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE BUR BREAKS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17306633 ↗MalfunctionJul 12, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE BURS THAT BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17305022 ↗MalfunctionJul 12, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE BURS THAT BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17305020 ↗MalfunctionJun 19, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE HEAD OF THE BUR BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17159421 ↗MalfunctionJun 15, 2023Fracture; Fracture
PER MEDWATCH REPORT#; MW5117493; IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT ALL PIECES OF THE BROKEN BUR WERE RETRIEVED SUCCESSFULLY. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17134154 ↗MalfunctionApr 13, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT ALL PARTS OF THE BROKEN BUR WERE RETRIEVED SUCCESSFULLY FROM THE PATIENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16731681 ↗MalfunctionMar 29, 2023Fracture; Overheating of Device; Fracture; Overheating of Device
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR HEAD OVERHEATED AND BROKE. IT WAS ALSO REPORTED THERE WAS A SHORT DELAY AND NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16639634 ↗MalfunctionMar 29, 2023Vibration; Vibration
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR WAS VIBRATING. IT WAS ALSO REPORTED THAT THE EVENT CAUSED A SHORT DELAY AND NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16635473 ↗MalfunctionSep 16, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE INSTANCE OF THE BURS BREAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15441876 ↗MalfunctionSep 14, 2022Fracture; Fracture
IT WAS REPORTED THAT PRIOR TO A SURGICAL PROCEDURE, IT WAS NOTED THAT THE BUR BROKE OUT OF THE PACKAGING. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15417491 ↗MalfunctionSep 14, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE BUR BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15417362 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page