Skip to content

N/A (REF 10-401FC)Recalls & reported problems

Stryker Sustainability Solutions, Inc. · MyoSure REACH Tissue Removal Device · Class 2

Quick summary

This Stryker MyoSure Reach tissue-removal device is used with a hysteroscope. The 26 reports mostly describe the device not working, jamming, showing torque errors, or shedding or separating parts; some mention uterine perforation. The reports include 9 injuries and no deaths. No recalls name this device, and reports fell from 14 in the previous 12 months to 5 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 07613327507133), received between Sep 1, 2023 and Sep 29, 2026 — classified as 8 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 3 reportsMay 2025: 1 reportsJun 2025: 4 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem10
  • Use of Device Problem5
  • Failure to Cycle4
  • Mechanical Jam3
  • Material Separation2
  • No Device Output2
  • Activation Problem1
  • Compatibility Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Insufficient Information4
  • Uterine Perforation2
  • Perforation1

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Activation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940822 ↗
  • MalfunctionSep 23, 2026Compatibility Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862955 ↗
  • InjuryMay 6, 2026Material Separation

    IT WAS REPORTED DURING THE CASE, THE BLADE CAME OFF AND WERE UNSURE IF THE PIECE WAS IN THE PATIENT OR IN THE SHAFT OF THE DEVICE. IT WAS FURTHER REPORTED AN XRAY WAS PERFORMED AND THERE WAS NO BROKEN PIECE FOUND. THERE WAS NO REPORTED PATIENT INJURY AND EXTENDED PROCEDURE TIME REPORTED WAS THEY NEEDED TO OPEN AN XL DEVICE AND CHANGE THE SCOPE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25106901 ↗
  • MalfunctionDec 19, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE MYOSURE OPERATED AS NORMAL AT THE BEGINNING OF THE CASE. THE PATIENT WAS HIGH IN THE BED AND THE DOCTOR HAD A HARD TIME VISUALIZING, SO THEY HAD TO STOP THE PROCEDURE AND MOVE THE PATIENT DOWN. UPON RESUMING THE END OF THE PROCEDURE, THE STAFF AND THE DOCTOR SAID THERE WAS ¿GLITTER¿/METAL SHAVINGS SWIRLING AROUND ON THE CAMERA AND THAT IT WAS FROM THE MYOSURE DEVICE." THE CIRCULATOR STATED THEY BELIEVED THE METAL SHAVINGS WERE FLUSHED OUT FROM THE FLUID MANAGEMENT SYSTEM. IT COULD NOT BE CONFIRMED SINCE THE SHAVINGS WERE SO SMALL. NO OTHER ACTION WAS TAKEN BY THE SURGEON OR THE TEAM FOR THE METAL SHAVINGS. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23857701 ↗
  • MalfunctionDec 2, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE MYOSURE DEVICE WAS DEFECTIVE. THE SURGEON REPORTED THAT THE MYOSURE REACH WAS SHEARING OFF METAL INSIDE THE PATIENT DURING USE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23698748 ↗
  • MalfunctionSep 24, 2025Failure to Cycle; Material Separation; Material Deformation; Failure to Cycle; Material Separation; Material Deformation

    IT WAS REPORTED THAT THE MYOSURE REACH WIRES AND CONNECTION FROM THE CORD TO THE HANDPIECE DISCONNECTED AND THE WIRES WERE EXPOSED. IT WAS DISCOVERED WHEN PLUGGED INTO THE FLUENT MANAGEMENT MACHINE AND WAS NOT USED ON THE PATIENT. AS REPORTED, THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23139871 ↗
  • MalfunctionAug 20, 2025Mechanical Problem; Decrease in Pressure; Device Displays Incorrect Message

    MYOSURE REACH ERROR "EXCESSIVE TORQUE DETECTED, REDUCE PRESSURE ON DEVICE, TAP FOOT PEDAL TO CONTINUE - (B)(4)." DEVICE WOULD NOT RELEASE FROM MACHINE. NEEDED TO OBTAIN NEW MACHINE AND HANDPIECE.

    Patient problems: Insufficient Information

    MAUDE report 22849167 ↗
  • InjuryJul 29, 2025Failure to Cycle; Failure to Cycle

    IT WAS REPORTED THAT THE MYOSURE WAS BEING USED AN GAVE AN ERROR MESSAGE OF "TOO MUCH TORQUE PRESS MEDAL" AND THE PEDAL WOULD NOT OPERATE. WHEN THE INSTRUMENT WAS PULLED OUT, IT LOOKED LIKE THE JAW WOULD NOT OPEN. THE SURGEON "DOESN'T THINK THE PATIENT HAD ANY HARM" BUT THE UTERUS WAS PERFORATED. AS REPORTED, THE PROCEDURE WAS CONVERTED TO A DIAGNOSTIC LAPAROSCOPY. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. NO OTHER INFORMATION WAS PROVIDED REGARDING MEDICAL INTERVENTION OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Uterine Perforation

    MAUDE report 22642731 ↗
  • MalfunctionJun 30, 2025Mechanical Problem; Difficult to Open or Close

    MYOSURE WAS BEING USED GAVE AN ERROR MESSAGE OF "TOO MUCH TORQUE PRESS PEDAL", PEDAL WOULDN'T OPERATE. WHEN PULLED INSTRUMENT OUT LOOKED LIKE JAW WOULDN'T OPEN. SURGEON "DOESN'T THINK THE PATIENT HAD ANY HARM" BUT THE UTERUS WAS PERFORATED. THE PROCEDURE WAS CONVERTED TO A DIAGNOSTIC LAPAROSCOPY. MANUFACTURER RESPONSE FOR HYSTEROSCOPE (AND ACCESSORIES), N/A (PER SITE REPORTER).

    Patient problems: Uterine Perforation

    MAUDE report 22352630 ↗
  • MalfunctionJun 25, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE MYOSURE WOULD NOT TURN ON/GIVE POWER. THE SYSTEM WAS REBOOTED AND STILL DID NOT WORK. AS REPORTED, OEM WAS PULLED AND WORKED FINE ON THE SAME GENERATOR. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS TEN MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332175 ↗
  • InjuryJun 19, 2025Failure to Cycle; Failure to Cycle

    IT WAS REPORTED THAT THERE WAS A FAULTY REPROCESSED MYOSURE HANDPIECE AND THE SURGEON WAS NOT HAPPY WITH THE DEVICE. ACCORDING TO THE SURGEON, IT IMPACTED THE PATIENT'S CARE AND THE PATIENT SCHEDULED FOR OUTPATIENT SURGERY HAD TO STAY THE NIGHT. THERE WERE NO DETAILS AVAILABLE ON WHY THE PATIENT WAS HAD TO THE STAY THE NIGHT. AS REPORTED, THE MYOSURE DEVICE WAS NOT CUTTING WELL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 15 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22294301 ↗
  • InjuryJun 19, 2025Failure to Cycle; Failure to Cycle

    IT WAS REPORTED THAT THERE WAS A FAULTY REPROCESSED MYOSURE HANDPIECE AND THE SURGEON WAS NOT HAPPY WITH THE DEVICE. ACCORDING TO THE SURGEON, IT IMPACTED THE PATIENT'S CARE AND THE PATIENT SCHEDULED FOR OUTPATIENT SURGERY HAD TO STAY THE NIGHT. THERE WERE NO DETAILS AVAILABLE ON WHY THE PATIENT HAD TO THE STAY THE NIGHT. AS REPORTED, THE MYOSURE DEVICE WAS NOT CUTTING WELL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 15 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22290112 ↗
  • MalfunctionMay 23, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A FAILURE OF THE MYOSURE ONCE THE SURGEON STARTED TO USE THE DEVICE. AS REPORTED, IT DID NOT WORK AFTER MULTIPLE TRIES. THE DEVICE WAS REPLACED WITH AN OEM DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS A FEW MINUTES TO GET OEM DEVICE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22080101 ↗
  • MalfunctionApr 25, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE MYOSURE DEVICE WOULD NOT ENGAGE THE BLADE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21912614 ↗
  • MalfunctionApr 3, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE MYOSURE DEVICE LOCKED UP. AS REPORTED, THE DEVICE WAS REPLACED WITH A STRYKER MYOSURE LITE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21767098 ↗
  • MalfunctionApr 3, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE MYOSURE CORD WOULD JUMP, ALMOST LIKE A SHORT IN THE WIRING AND THEN THE EQUIPMENT WOULD SHUT OFF. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21763520 ↗
  • MalfunctionMar 13, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE MYOSURE DEVICE IS NOT CUTTING AND MAKING GRINDING NOISE DURING SURGICAL PROCEDURE. DEVICE WAS REPLACED. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21599452 ↗
  • MalfunctionFeb 5, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE MYOSURE DEVICE WAS NOT CUTTING. AS REPORTED, IT WAS REPLACED WITH ANOTHER STRYKER DEVICE AND WORKED FINE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED AND THE EXTENDED PROCEDURE TIME REPORTED WAS FIVE MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21313255 ↗
  • MalfunctionJan 14, 2025Defective Component; Defective Component

    IT WAS REPORTED THAT THE MYOSURE DEVICE SOUNDED LIKE GRINDING METAL, WAS NOT TURNING PROPERLY, AND WAS DIFFICULT TO REMOVE FROM THE MACHINE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21151833 ↗
  • MalfunctionOct 3, 2024No Device Output; No Device Output

    IT WAS REPORTED THAT A MYOSURE DEVICE WAS JUMPING AND READING (REVVING) UP THE TORQUE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS MINIMAL. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20362248 ↗
  • MalfunctionJun 1, 2024No Device Output; No Device Output

    IT WAS REPORTED THAT WHEN THE MYOSURE WAS PLUGGED IN AND ACTIVATED, THE SHAVER PART STARTED AND THEN STOPPED WORKING ALMOST IMMEDIATELY AFTER IT STARTED. IT WOULD TURN ON AND OFF A COUPLE TIMES, THEN NOT WORK AT ALL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS THREE MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19442097 ↗
  • MalfunctionMay 16, 2024Mechanical Problem; Use of Device Problem; Mechanical Problem; Use of Device Problem

    IT WAS REPORTED THE DEVICE BLADES FAILED TO ACTUATE. NO INFORMATION WAS PROVIDED REGARDING PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19336084 ↗
  • InjuryApr 22, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DEVICE WOULD NOT WORK CONSISTENTLY AND PREVENTED THE DOCTOR FROM PERFORMING SURGERY AS PLANNED. HAD TO USE ANOTHER DEVICE AND DIFFERENT PIECE OF EQUIPMENT. COMPLAINANT NOT AWARE OF ANY OTHER PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19160615 ↗
  • InjuryApr 22, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DEVICE WOULD NOT WORK CONSISTENTLY AND PREVENTED THE DOCTOR FROM PERFORMING SURGERY AS PLANNED. HAD TO USE ANOTHER DEVICE AND DIFFERENT PIECE OF EQUIPMENT. DEVICE CABLE HAD TO BE PLIED OUT OF FLUENT MANAGEMENT SYSTEM. COMPLAINANT NOT AWARE OF ANY OTHER PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19160613 ↗
  • InjuryNov 7, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THE DEVICE WAS COMPLETED WITH DILATION AND CURETTAGE (D&C) BECAUSE THE COMPLAINT DEVICE ACTIVATED AND FIRED BUT THEN FROZE. THERE WAS NO OTHER PATIENT INJURY OR MEDICAL INTERVENTION REPORTED AND THE EXTENDED PROCEDURE TIME REPORTED WAS FIVE MINUTES TO REPLACE THE DEVICE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18086846 ↗

Other HYSTEROSCOPE (AND ACCESSORIES) devices

All HYSTEROSCOPE (AND ACCESSORIES) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.