Skip to content

iBur 8431-107-030DRecalls & reported problems

Stryker Corporation · 3.0mm DIAMOND MATCH HEAD, DISTAL BEND · Class 2

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327509229), received between Dec 23, 2021 and Jan 23, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Fracture” (6 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 24, 2024Completed

    Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

    Z-2730-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Fracture6
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163993 ↗
  • MalfunctionOct 28, 2025Fracture; Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FRACTURE. 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399924 ↗
  • MalfunctionJul 8, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING AN THE TIP OF THE BUR WAS BURNT DURING AN ENDOSCOPIC TRANSNASAL PITUITARY TUMOR RESECTION THE TIP OVERHEATED AND BECAME BURNT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19694964 ↗
  • MalfunctionNov 22, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE DURING USE. IT WAS ALSO REPORTED THE EVENT CAUSED A 90 MINUTE DELAY TO THE PROCEDURE. IT WAS FURTHER REPORTED THAT THERE WAS NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES THE SECOND INSTANCE OF THE BUR BREAK WHICH WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18192599 ↗
  • MalfunctionNov 22, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE DURING USE. IT WAS ALSO REPORTED THE EVENT CAUSED A 90 MINUTE DELAY TO THE PROCEDURE. IT WAS FURTHER REPORTED THAT THERE WAS NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18192598 ↗
  • MalfunctionJul 12, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP OF THE BUR BROKE. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE EVENT CAUSED A DELAY OF LESS THAN 5 MINUTES. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17304939 ↗
  • MalfunctionDec 23, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE BUR BROKE . IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13076948 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.