iBur 8431-107-030DRecalls & reported problems
Stryker Corporation · 3.0mm DIAMOND MATCH HEAD, DISTAL BEND · Class 2
All models of iBur (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327509229), received between Dec 23, 2021 and Jan 23, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (6 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 24, 2024Class IICompleted
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Z-2730-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture6
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163993 ↗MalfunctionOct 28, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FRACTURE. 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399924 ↗MalfunctionJul 8, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING AN THE TIP OF THE BUR WAS BURNT DURING AN ENDOSCOPIC TRANSNASAL PITUITARY TUMOR RESECTION THE TIP OVERHEATED AND BECAME BURNT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19694964 ↗MalfunctionNov 22, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE DURING USE. IT WAS ALSO REPORTED THE EVENT CAUSED A 90 MINUTE DELAY TO THE PROCEDURE. IT WAS FURTHER REPORTED THAT THERE WAS NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES THE SECOND INSTANCE OF THE BUR BREAK WHICH WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18192599 ↗MalfunctionNov 22, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, TWO BURS BROKE DURING USE. IT WAS ALSO REPORTED THE EVENT CAUSED A 90 MINUTE DELAY TO THE PROCEDURE. IT WAS FURTHER REPORTED THAT THERE WAS NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18192598 ↗MalfunctionJul 12, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP OF THE BUR BROKE. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE EVENT CAUSED A DELAY OF LESS THAN 5 MINUTES. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17304939 ↗MalfunctionDec 23, 2021Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE BUR BROKE . IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13076948 ↗
FDA clearance and classification
The FDA regulates product code HBE (drills, burrs, trephines & accessories (simple, powered)) under 21 CFR 882.4310 as a Class II device. 21 CFR 882.4310 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the iBur 8431-107-030D
Has the iBur 8431-107-030D been recalled?
Yes. 1 FDA recall record names this device's identifier, and the company has completed it but the FDA has not yet closed it. The recall, Z-2730-2024 (Class II), began on Jul 24, 2024: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the iBur 8431-107-030D?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Dec 23, 2021 and Jan 23, 2026. In those reports, the most common specific device problems are “Fracture” (6) and “Overheating of Device” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the iBur 8431-107-030D?
None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the iBur 8431-107-030D, and is it still sold?
The FDA's device database lists Stryker Corporation as the company responsible for it (brand iBur, model 8431-107-030D). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,624 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)930 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)414 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)221 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)186 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)184 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)182 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)164 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice