Skip to content

GAMMA 1420-0240Recalls & reported problems

Stryker GmbH · Lag Screw Reamer · Class 1

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327510133), received between Oct 27, 2025 and Jun 18, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align3
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2026Failure to Align

    THE SALES REP REPORTED THAT WHEN IN USE, THE REAMER OSCILLATES ABNORMALLY. ADDITIONAL INFORMATION RECEIVED: THE PROCEDURE WAS COMPLETED, BUT THE OPERATING TIME WAS EXTENDED. THIS HAD AN IMPACT ON THE PATIENT, AS THE DRILL BIT RUBBED AGAINST THE NAIL, CAUSING TITANIUM DEBRIS TO ENTER THE PATIENT¿S BODY. TWO MOTORS WERE USED TO VERIFY THAT THE PROBLEM WAS NOT WITH THE DRILL CHUCK, BUT THE ISSUE PERSISTED. FOR THE PATIENT, THERE WAS AN INCREASED RISK OF INFECTION DUE TO THE LONGER PROCEDURE TIME AND MORE FREQUENT ENTRIES AND EXITS FROM THE OPERATING ROOM. INCREASED STERILIZATION VOLUME, LONGER OPERATING ROOM OCCUPANCY.

    Patient problems: Foreign Body In Patient

    MAUDE report 25548587 ↗
  • MalfunctionMar 23, 2026Material Fragmentation; Failure to Align

    AS REPORTED: "WHEN USING THE LAG SCREW DRILL, THERE WAS SLIGHT RESISTANCE, AND WHEN REMOVING THE DRILL BIT, A METAL-LIKE FRAGMENT WAS ADHERING TO IT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24661194 ↗
  • MalfunctionOct 27, 2025Failure to Align; Failure to Align

    AS REPORTED: "CUSTOMER WAS NOT ABLE TO HIT THE NAIL CORRECTLY WITH THE FEMORAL NECK SCREW DRILL. THE NAIL WAS HIT WITH THE DRILL, THIS ALSO HAPPENED WITH THE SECOND NAIL. THE NAIL HOLDING SCREWS SHOW SIGNS OF WEAR." ADDITIONAL INFORMATION: ANOTHER NAIL WAS IMPLANTED AFTER THE FIRST NAIL WAS REMOVED. THE RESULT WITH THE SECOND NAIL WAS GOOD; THE DRILLED NAIL WAS NOT LEFT IN SITU.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23389611 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page