Stryker Corporation · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker 1788 camera control unit is the video processing tower used with a laparoscope in general and plastic surgery. The 31 adverse event reports mostly describe the video going black or cutting out mid-procedure, sometimes shutting down entirely and needing a replacement tower to finish the case. No deaths and no injuries were reported, and all 31 reports are malfunctions. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 07613327580013), received between Oct 11, 2023 and Sep 25, 2026 — classified as 38 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 5 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 6 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image32
  • Connection Problem10
  • Poor Quality Image4
  • Erratic or Intermittent Display2
  • Optical Problem2
  • Device Displays Incorrect Message1
  • Disconnection1
  • Display or Visual Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions36
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026No Display/Image

    It was reported that flickering image was observed during the case. Therefore, there may have been a full loss of image.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898257 ↗
  • MalfunctionSep 24, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888860 ↗
  • MalfunctionSep 23, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873505 ↗
  • MalfunctionSep 11, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50744446 ↗
  • MalfunctionSep 2, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50564921 ↗
  • MalfunctionSep 1, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50549689 ↗
  • MalfunctionJul 9, 2026No Display/Image; Power Problem

    IT WAS REPORTED THAT THE DEVICE COULDN'T ROUTE IN THROUGH THE MONITOR. THEREFORE THERE WAS NO IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25771852 ↗
  • MalfunctionMar 26, 2026Erratic or Intermittent Display

    IT WAS REPORTED THAT THERE WAS A ROLLING BLACK BAR ON THE TOP OF THE MONITOR. THERE WAS NO FULL LOSS OF IMAGE, BUT PART OF THE LIVE IMAGE WAS CUT OFF, CAUSING A PARTIAL IMAGE LOSS. A REPLACEMENT TOWER WAS USED TO COMPLETE THE MEDICAL PROCEDURE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24702052 ↗
  • MalfunctionJan 12, 2026No Display/Image; Failure to Power Up

    IT WAS REPORTED THAT THE CCU SHUT DOWN IN THE MIDDLE OF A CASE COMPLETELY UNPROMPTED AND NOW WILL NOT TURN BACK ON. THE CCU WAS REMOVED AND SWAPPED OUT WITH ANOTHER FROM A TOWER AT THE FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24046497 ↗
  • MalfunctionDec 29, 2025No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS REPORTED THAT THE DEVICE TURNED OFF DURING THE MIDDLE OF THE CASE, AND THE PROCEDURE WAS COMPLETED USING A REPLACEMENT DEVICE. THE FULL LOSS OF IMAGE DURATION COULD NOT BE CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23916281 ↗
  • MalfunctionJul 10, 2025No Display/Image; No Display/Image

    THE UNIT FLICKERED MID CASE BACK TO THE "PLUG CAMERA IN" SCREEN.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22472939 ↗
  • MalfunctionJun 19, 2025No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS REPORTED BY THE CUSTOMER THAT THE 1788 CCU UNIT WAS NOT WORKING SMOOTHLY, WITH THE CAMERA GOING BLACK UNPREDICTABLY. THE STRYKER REP FURTHER REPORTED THAT HE TRIED TO TROUBLESHOOT THE UNIT, BUT THE ISSUE STILL OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22294263 ↗
  • MalfunctionApr 3, 2025No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21766988 ↗
  • MalfunctionJan 14, 2025No Display/Image; Failure to Sense; No Display/Image; Failure to Sense

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21153791 ↗
  • MalfunctionJan 9, 2025No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21112643 ↗
  • MalfunctionJan 2, 2025No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21061224 ↗
  • MalfunctionDec 18, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: Insufficient Information

    MAUDE report 20970903 ↗
  • MalfunctionDec 18, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: Insufficient Information

    MAUDE report 20970491 ↗
  • MalfunctionNov 27, 2024No Display/Image; Output Problem; No Display/Image; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20797832 ↗
  • MalfunctionNov 27, 2024No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20790046 ↗
  • MalfunctionAug 28, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20091846 ↗
  • MalfunctionAug 22, 2024Disconnection; No Display/Image; Disconnection; No Display/Image

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20042507 ↗
  • MalfunctionJul 3, 2024No Display/Image; Display or Visual Feedback Problem; No Display/Image; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19671522 ↗
  • MalfunctionJul 3, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19671431 ↗
  • MalfunctionJul 3, 2024No Display/Image; No Display/Image

    THIS EVENT WAS INITIALLY REPORTED AS A MALFUNCTION FOR LOSS OF IMAGE. UPON ADDITIONAL INFORMATION PROVIDED, IT WAS CONFIRMED THAT A LOSS OF IMAGE DID NOT OCCUR. THEREFORE, THIS EVENT IS NO LONGER REPORTABLE. THIS EVENT DESCRIPTION MOST LIKELY INDICATES "CONFUSING ERRORS OR NOTIFICATIONS" AND ¿NO IMAGE¿. THE REVIEW OF HISTORICAL COMPLAINT DATA, RISK DOCUMENTATION, AND THE REPORTED INFORMATION REGARDING THIS EVENT DO NOT SUGGEST A RECURRENCE OF THIS EVENT WOULD BE LIKELY TO RESULT IN A SERIOUS INJURY. NO ADVERSE CONSEQUENCES WERE REPORTED, IT DOES NOT SEEM LIKELY THAT ADVERSE CONSEQUENCES WOULD RESULT IF IT WERE TO RECUR. NO REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19671429 ↗

Other Laparoscope, general & plastic surgery devices