1788 1788610122Recalls & reported problems

Stryker Corporation · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker laparoscope is a rigid scope used in general and plastic surgery. The 30 reports mostly describe the camera producing no picture or image at all, along with flickering, poor quality images, connection problems, and loss of signal during procedures. No deaths or injuries are reported. There are no recalls for this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 07613327580044), received between Jan 4, 2024 and Sep 22, 2026 — classified as 35 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (29 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 9 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image29
  • Output Problem6
  • Poor Quality Image5
  • Connection Problem3
  • Erratic or Intermittent Display2
  • Temperature Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions31
  • Insufficient Information6

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862667 ↗
  • MalfunctionSep 18, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50831245 ↗
  • MalfunctionSep 14, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764849 ↗
  • MalfunctionSep 3, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50576877 ↗
  • MalfunctionSep 1, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50549693 ↗
  • MalfunctionAug 31, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50534176 ↗
  • MalfunctionAug 31, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50532322 ↗
  • MalfunctionAug 31, 2026No Display/Image

    It was reported that there was picture issues/flickering with the camera.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50532142 ↗
  • MalfunctionAug 31, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50532141 ↗
  • MalfunctionAug 31, 2026Poor Quality Image

    It was reported that there was bad image.

    Patient problems: Insufficient Information

    MAUDE report 50526583 ↗
  • MalfunctionAug 12, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50229355 ↗
  • MalfunctionAug 5, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50149872 ↗
  • MalfunctionAug 5, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50149869 ↗
  • MalfunctionAug 3, 2026No Display/Image

    It was reported that during the case the camera would intermittently drop out and then restart. Therefore, there was loss of image.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50124133 ↗
  • MalfunctionMay 15, 2026No Display/Image; Output Problem

    IT WAS REPORTED BY THE CUSTOMER THAT THE CAMERA HEAD WAS FLICKERING. THE DURATION OF LOSS OF IMAGE COULD NOT BE CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25192859 ↗
  • MalfunctionMay 15, 2026No Display/Image; Output Problem

    IT WAS REPORTED BY THE CUSTOMER THAT THE CAMERA HEAD WAS FLICKERING. HOWEVER, THE DURATION COULD NOT BE CONFIRMED, THEREFORE THERE WAS LOSS OF IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25192094 ↗
  • MalfunctionApr 2, 2026No Display/Image; Output Problem

    IT WAS REPORTED THAT IT WAS REPORTED THAT THE UNIT OCCASIONALLY ACTS UP, EITHER LOSING CONNECTION TO THE CCU OR FAILING THE WHITE BALANCE. THERE WAS A FULL LOSS OF IMAGE FOR A FEW MINUTES WHILE THE THE CAMERA HEAD WAS SWAPPED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765199 ↗
  • MalfunctionApr 1, 2026No Display/Image; Output Problem

    IT WAS REPORTED THAT IT WAS REPORTED THAT THE UNIT OCCASIONALLY ACTS UP, EITHER LOSING CONNECTION TO THE CCU OR FAILING THE WHITE BALANCE. THERE WAS A FULL LOSS OF IMAGE FOR A FEW MINUTES WHILE THE CAMERA HEAD WAS SWAPPED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24759399 ↗
  • MalfunctionMar 19, 2026No Display/Image

    IT WAS REPORTED THAT THE CAMERA THAT IS CUTTING OUT DURING SURGERY. THERE WAS FULL LOSS OF IMAGE AND THIS CAUSED A 5 MINUTE DELAY. A NEW DEVICE WAS USED TO COMPLETE THE MEDICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24641427 ↗
  • MalfunctionDec 22, 2025Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image

    IT WAS REPORTED BY THE CUSTOMER THAT THE CAMERA WAS FLICKERING WHEN THE CABLE IS MOVED. THERE WAS PARTIAL IMAGE LOSS, REPLACEMENT DEVICE USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872866 ↗
  • MalfunctionDec 19, 2025Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image

    IT WAS REPORTED BY THE CUSTOMER THAT THE IMAGE FLICKERS WHEN THE CABLE IS MOVED. THERE WAS PARTIAL IMAGE LOSS, REPLACEMENT DEVICE USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23855965 ↗
  • MalfunctionDec 18, 2025No Display/Image; Output Problem; No Display/Image; Output Problem

    IT WAS REPORTED THAT THE CAMERA HEAD IS DISCONNECTING WHEN THE CABLE IS MOVED NEAR THE CAMERA HEAD SIDE. THEREFORE THERE WAS LOSS OF IMAGE.

    Patient problems: Insufficient Information

    MAUDE report 23843467 ↗
  • MalfunctionJul 11, 2025No Display/Image; No Display/Image

    THE UNIT FLICKERED MID CASE BACK TO THE "PLUG CAMERA IN" SCREEN. THEREFORE, THERE WAS LOSS OF IMAGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22479973 ↗
  • MalfunctionJun 24, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE SCREEN INTERMITTENTLY LOST CONNECTION FOR 10¿20 SECONDS BEFORE COMING BACK ONLINE. EACH TIME THE CONNECTION COMES BACK, WHITE BALANCING WAS REQUIRED BEFORE THE PROCEDURE COULD CONTINUE. ADDITIONALLY, THE LIGHT WOULD INTERMITTENTLY CHANGE IN INTENSITY AND COLOR, ALSO REQUIRING WHITE BALANCING. INITIAL ACTIONS TAKEN INCLUDED FIXING, DISCONNECTING, AND RECONNECTING THE CAMERA HEAD AND HDMI PORT, AS WELL AS RESTARTING THE SYSTEM. THIS FIXED THE ISSUE TEMPORARILY. WHEN THE ISSUE RECURRED, THE CAMERA HEAD WAS REPLACED, WHICH APPEARED TO RESOLVE THE ISSUE FOR THE REMAINDER OF THE PROCEDURE. NO SURGICAL DELAY REPORTED OTHER THAN A FEW MINUTES TO SWAP THE CAMERA HEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22323605 ↗
  • MalfunctionApr 29, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE CAMERA HEAD HAS BEEN GOING IN AND OUT OF CONNECTION AND WHEN IT IS BENT A SPECIFIC WAY, IT LOSES CONNECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21928866 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates