Color Cuff 5921-218-135NSRecalls & reported problems
Stryker Corporation · Nonsterile Disposable Tourniquet Cuff, Single Bladder, Single Port and Quick Connect Connector · Class 1
All models of Color Cuff (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327606133), received between Mar 7, 2023 and Dec 22, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Decrease in Pressure” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 1, 2023Class IIOpen, Classified
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications
Z-1343-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Decrease in Pressure5
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2023Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE DEVICE INFLATED AND THEN DEFLATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18380911 ↗MalfunctionDec 21, 2023Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE DEVICE INFLATED AND THEN DEFLATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18371747 ↗MalfunctionDec 21, 2023Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE DEVICE INFLATED AND THEN DEFLATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18371745 ↗MalfunctionApr 4, 2023Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT CUFF LEAKED AIR DURING USE IN A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16676286 ↗MalfunctionMar 7, 2023Decrease in Pressure; Decrease in Pressure
THIS REPORT SUMMARIZES 50 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 34 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 15 EVENTS HAD INSUFFICIENT INFORMATION RECEIVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16500408 ↗
Other TOURNIQUET, PNEUMATIC devices
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall82 reports
- Color Cuff 5921018135Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls31 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall72 reports
- Color Cuff 5921018235Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls20 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall38 reports
- SmartPump 5920011000Stryker Corporation · TOURNIQUET, PNEUMATIC72 reports
- A.T.S.® Personalized Pressure® 60320010100Zimmer Orthopaedic Surgical Products, Inc. · TOURNIQUET, PNEUMATIC1 recall20 reports
- Color Cuff 5921-034-135NSStryker Corporation · TOURNIQUET, PNEUMATIC1 recall20 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page