Skip to content

Color Cuff 5921-218-135NSRecalls & reported problems

Stryker Corporation · Nonsterile Disposable Tourniquet Cuff, Single Bladder, Single Port and Quick Connect Connector · Class 1

All models of Color Cuff (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327606133), received between Mar 7, 2023 and Dec 22, 2023 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Decrease in Pressure” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Class IIOpen, Classified

    Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications

    Z-1343-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Decrease in Pressure5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2023Decrease in Pressure; Decrease in Pressure

    IT WAS REPORTED THAT THE DEVICE INFLATED AND THEN DEFLATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18380911 ↗
  • MalfunctionDec 21, 2023Decrease in Pressure; Decrease in Pressure

    IT WAS REPORTED THAT THE DEVICE INFLATED AND THEN DEFLATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18371747 ↗
  • MalfunctionDec 21, 2023Decrease in Pressure; Decrease in Pressure

    IT WAS REPORTED THAT THE DEVICE INFLATED AND THEN DEFLATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18371745 ↗
  • MalfunctionApr 4, 2023Decrease in Pressure; Decrease in Pressure

    IT WAS REPORTED THAT CUFF LEAKED AIR DURING USE IN A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16676286 ↗
  • MalfunctionMar 7, 2023Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 50 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 34 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 15 EVENTS HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16500408 ↗

Other TOURNIQUET, PNEUMATIC devices

All TOURNIQUET, PNEUMATIC devices →

Sources and updates