Skip to content

PANGEA 541037Recalls & reported problems

Stryker GmbH · 3.5 Straight Narrow Plate · Class 2 · Implant

All models of PANGEA (53)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327606522), received between Nov 24, 2025 and Nov 24, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Difficult to Setup or Prepare1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    AS REPORTED: "THE PRODUCTS IN QUESTION ARE THE 3.5MM NARROW COMPRESSION PLATE T15, (541037 7 HOLE PLATE FOR EXAMPLE) AND THE SHORTER T15 CORTICAL SCREWS (661412 12MM SCREW FOR EXAMPLE). I WATCHED THE SURGEON DRILL THROUGH BOTH CORTICES ONCE USING THE VA GUIDE (542028) IN WHAT LOOKED LIKE A PERPENDICULAR TO THE PLATE TRAJECTORY. WE CONFIRMED IT WAS A 2.5MM DRILL USED. THE SCREW WOULD CONTINUE TO SPIN IN THE HOLE UPON INSERTION, THIS HAPPENED TWICE. WE THEN USED A LONGER SCREW AND THE SAME HAPPENED AGAIN. THE SURGEON THINKS IT'S BECAUSE THE PITCH OF THE SCREWS ARE INFERIOR AND HENCE ON SMALLER LENGTHS THEY DON'T GET AS GOOD AS PURCHASE AS WHAT HE USES. ANOTHER COMMENT WAS THAT THE SCREW HEAD LOOKED TOO SMALL COMPARED TO THE DCP HOLE. THIS WAS A PROBLEM WHEN A SCREW WAS INSERTED ON AN ANGLE AWAY FROM THE FRACTURE SITE. THE SURGEON FOUND THAT UPON TIGHTENING THE CORTICAL SCREW IN THIS TRAJECTORY THAT THE SPACE BETWEEN THE SCREW HEAD AND THE BOTTOM OF THE PLATE WAS VERY LOW AND HE WAS WORRIED ABOUT SCREW PULL THROUGH ON THE PLATE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23639301 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page