The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613336154304), received between Sep 19, 2025 and Sep 19, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result3
- Incorrect, Inadequate or Imprecise Result or Readings2
- Incorrect Measurement1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 19, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER ALLEGED A DISCREPANT REACTIVE RESULT FOR SAMPLE VM-10 FROM AN API SURVEY USING THE ANTI-HCV G2 ELECSYS ASSAY ON THE COBAS 6000 E601 MODULE. THE RESULT WAS EXPECTED TO BE NON-REACTIVE BUT RECOVERED AS REACTIVE. THE CUSTOMER RE-RAN THE API SURVEY MATERIAL, AND THE SAME FALSE POSITIVE RESULT WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23094360 ↗MalfunctionSep 19, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER ALLEGED A FAILURE OF THE ANTI-HCV G2 ELECSYS COBAS E 100 ASSAY ON THE COBAS 6000 E601 MODULE DURING AN API SURVEY. THE CUSTOMER REPORTED THAT SAMPLE (B)(6), WHICH WAS EXPECTED TO RESULT AS NONREACTIVE, INSTEAD RESULTED AS REACTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23094359 ↗MalfunctionSep 19, 2025False Positive Result; Incorrect Measurement; False Positive Result; Incorrect Measurement
THE INITIAL REPORTER ALLEGED A DISCREPANT REACTIVE RESULT FOR A SAMPLE (VM-10) TESTED USING THE ANTI-HCV G2 ELECSYS ASSAY ON THE COBAS 6000 E601 MODULE. THE SAMPLE WAS PART OF AN API SURVEY, WHERE THE EXPECTED RESULT WAS NON-REACTIVE. THE INITIAL TEST RESULT FOR THE SAMPLE WAS 1.40 COI (REACTIVE). THE SAMPLE WAS MANUALLY LOADED INTO A PRIMARY TUBE. ON (B)(6) 2020, THE SAMPLE WAS RETESTED ON THE SAME SYSTEM, YIELDING A RESULT OF 1.45 COI (REACTIVE). THE SAMPLE WAS THEN SENT TO A SISTER HOSPITAL, WHERE IT WAS TESTED AGAIN AND REMAINED REACTIVE. TO VERIFY THE ANALYZER'S PERFORMANCE, THE CUSTOMER RERAN A KNOWN NEGATIVE SAMPLE, WHICH PRODUCED A NON-REACTIVE RESULT. QUALITY CONTROL (QC) WAS REPORTED TO BE WITHIN SPECIFICATIONS. THE CUSTOMER SUSPECTED A SAMPLE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23094356 ↗
Other Assay,Enzyme Linked Immunosorbent,Hepatitis C Virus devices
- ADVIA Centaur® HCV 10309061Siemens Healthcare Diagnostics Inc. · Assay,Enzyme Linked Immunosorbent,Hepatitis C Virus108 reports
- ARCHITECT 1L79-25Abbott GmbH · ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS1 recall7 reports
- ARCHITECT 1L79-35Abbott GmbH · ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS1 recall3 reports
- Alinity 08P0521Abbott GmbH · Assay, enzyme linked immunosorbent, hepatitis c virus1 recall2 reports
- Elecsys Anti-HCV II 08837058190Roche Diagnostics GmbH · Assay, enzyme linked immunosorbent, hepatitis c virus44 reports
- Elecsys Anti-HCV II 08836981190Roche Diagnostics GmbH · Assay, enzyme linked immunosorbent, hepatitis c virus31 reports
- Elecsys Anti-HCV II 07026889190Roche Diagnostics GmbH · ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS26 reports
- Atellica™ IM aHCV 10995456Siemens Healthcare Diagnostics Inc. · ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page