Elecsys HBsAg IIRecalls & reported problems across all models
Roche Diagnostics GmbH · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 105
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 0
We found 5 models of Elecsys HBsAg II by Roche Diagnostics GmbH in the FDA's device registry (GUDID). Together they are named in 105 adverse event reports in the FDA's MAUDE database, received between Aug 13, 2021 and Sep 29, 2026 — classified as 105 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (68 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Elecsys HBsAg II problems
Device problems
- False Positive Result68
- Non Reproducible Results35
- False Negative Result17
- Incorrect, Inadequate or Imprecise Result or Readings7
- Contamination /Decontamination Problem2
- Improper or Incorrect Procedure or Method2
- Low Test Results2
- Output Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions104
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Hepatitis1
- Insufficient Information1
Elecsys HBsAg II models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Elecsys HBsAg II 08814848190Test, hepatitis b (b core, be antigen, be antibody, b core igm)93 reports
- Elecsys HBsAg II 08814848162Serology tests used for infectious disease screening of blood and plasma donations5 reports
- Elecsys HBsAg II 07251076190Test, hepatitis b (b core, be antigen, be antibody, b core igm)4 reports
- Elecsys HBsAg II 08814856160Test, hepatitis b (b core, be antigen, be antibody, b core igm)2 reports
- Elecsys HBsAg II 08814864160Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 report
Recent Elecsys HBsAg II adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937690 ↗MalfunctionSep 28, 2026False Positive Result
There was an allegation of questionable Elecsys HBsAg II results for 1 patient sample on a cobas e 801 analytical unit compared to an Abbott analyzer.The initial result was 0.9 COI (borderline). The repeat result was 0.945 COI.The sample was repeated on an Abbott analyzer and the result was 0.0 IU/mL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917598 ↗MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915516 ↗MalfunctionSep 23, 2026False Positive Result
There was an allegation of questionable results with the Elecsys HBsAg II assay for a donor sample tested on the Roche instrument (details have been requested). It was reported that the patient has always been non-reactive on a non-Roche instrument. Since switching to Roche instruments, the last two times the patient has been reactive for HBsAg but then confirmed negative for HBsAg.The test was repeated as the reactive result did not match previous history.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870259 ↗MalfunctionSep 22, 2026False Positive Result; Non Reproducible Results
The initial reporter questioned a positive result for 1 patient sample tested for Elecsys HBsAg II (HBsAg II) on a cobas e 801 analytical unit. The initial result was 4.97 COI (positive).The repeat results were 0.338 COI (negative) and 0.365 COI (negative). The sample was repeated, and the result was 0.359 COI (negative).The sample was repeated on the customer¿s other cobas pro analyzer, and the results were 0.399 COI (negative) and 0.326 COI (negative).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50857207 ↗MalfunctionSep 21, 2026False Positive Result; Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50846206 ↗MalfunctionSep 10, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50718093 ↗MalfunctionSep 10, 2026Incorrect, Inadequate or Imprecise Result or Readings
There was an allegation of questionable Elecsys HBsAg II results from the cobas e 411 disk analyzer.The customer stated that most of the HBsAg samples show low positive /borderline results at approximately 1 COI. This issue occurred repeatedly in multiple samples. No specific data or repeat testing results were provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713961 ↗MalfunctionAug 31, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50518999 ↗MalfunctionAug 17, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50286936 ↗MalfunctionAug 17, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50272184 ↗MalfunctionAug 14, 2026False Negative Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50258555 ↗MalfunctionAug 13, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50233745 ↗MalfunctionAug 5, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50139204 ↗MalfunctionJul 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50086585 ↗
Other Roche Diagnostics GmbH products
- cobas e 411 Immunoassay Analyzer2 models · 1,218 reports
- cobas e 801 Immunoassay analyzer3 models · 713 reports · 1 recall
- CoaguChek XS PT strips2 models · 632 reports
- CoaguChek XS Patient Self Testing Care Kit2 models · 346 reports
- Elecsys FT3 III4 models · 283 reports
- Elecsys Estradiol III2 models · 162 reports
- Elecsys TSH4 models · 155 reports
- Sodium electrode3 models · 150 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page