Skip to content

Elecsys HBsAg II 08814856160Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Elecsys HBsAg II (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336198285), received between Sep 15, 2025 and Aug 5, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2026False Positive Result

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50139204 ↗
  • MalfunctionSep 15, 2025False Positive Result; False Positive Result

    THE INITIAL REPORTER ALLEGED HIGH PATIENT RECOVERY AND FALSE-POSITIVE RESULTS WITH THE HBSAG G2 ELECSYS COBAS E 100 V2 ASSAY ON THE COBAS E411 DISK. THE FIRST INSTANCE INVOLVED A PATIENT SAMPLE TESTED ON (B)(6) 2021. THE INITIAL RESULT WAS 1.92 (REACTIVE). A FIRST RERUN ON THE PRIMARY TUBE YIELDED 0.308 (NON-REACTIVE), AND A SECOND RERUN FROM A HITACHI CUP YIELDED 0.310 (NON-REACTIVE). THE SECOND INSTANCE INVOLVED ANOTHER PATIENT SAMPLE DRAWN AND TESTED ON (B)(6) 2021. THE INITIAL RESULT WAS 1.24 (REACTIVE). A FIRST RERUN ON THE PRIMARY TUBE YIELDED 0.355 (NON-REACTIVE), AND A SECOND RERUN YIELDED 0.332 (NON-REACTIVE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23058000 ↗

Other Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices

All Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page