HAMILTON-C1 1610010Recalls & reported problems
Hamilton Medical AG · Class 2
All models of HAMILTON-C1 (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Hamilton-C1 ventilator has 167 reports, including 3 deaths and 4 injuries. Reports describe the ventilator losing power, failing to deliver full breaths or flow, leaks, and alarms during patient use. The device has 1 recall, still open, from July 2023, over control board capacitors that can leak and interrupt breathing support.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 163
- Deaths reported
- 3
- Injuries reported
- 6
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 163 adverse event reports that list this exact device (DI 07630002813426), received between Sep 2, 2022 and Sep 28, 2026 — classified as 3 deaths, 6 injuries, 154 malfunctions.
The most frequently reported specific device problem is “Complete Loss of Power” (22 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 26, 2023Open, Classified
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Z-2525-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Complete Loss of Power22
- Gas/Air Leak17
- Insufficient Device Problem Information17
- Failure to Deliver13
- No Flow13
- Defective Component9
- Failure of Device to Self-Test9
- Gas Output Problem7
Patient problems
- No Clinical Signs, Symptoms or Conditions153
- Insufficient Information8
- Low Oxygen Saturation2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Hypoxia1
- Respiratory Arrest1
- Respiratory Insufficiency1
Recent adverse event reports
The 25 most recent of 163 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026No Audible Alarm; No Apparent Adverse Event
Hamilton Medial AG received the following complaint description (summary):It was reported that, on (b)(6) 2026, a HAMILTON-C1 (Product No. 1610010, S/N (b)(6), SW 3.1.1) "did not audibly alarm while alarm condition existed on patient." Patient involvement was reported. No patient, user, or third-party harm and no medical intervention were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913717 ↗InjuryAug 7, 2026Adverse Event Without Identified Device or Use Problem
RT was called to bedside due to nursing stating that ventilator was spontaneously shutting off. RT came to bedside and nursing stated it happened while suctioning. RT suctioned patient and all flow from the ventilator stopped completely and patient O2 saturations dropped to low 80's and required support from AMBU bag to improve saturations. RT replaced flow sensor and exhalation valve and still the machine shut off all flow after suctioning despite the patient spontaneously breathing.
Patient problems: Low Oxygen Saturation
MAUDE report 50182653 ↗InjuryAug 7, 2026Insufficient Device Problem Information
While switching oxygen from tank to wall, alarm appeared "No oxygen " or "oxygen supply failed ". Duplicated the event in department and noticed that it took some time for the C1 to alarm that the vent had no oxygen. Once the O2 hose was plugged back in, the vent took some time to stop alarming "no oxygen. " It seems to reset itself because the alarm disappears. Unsure if this is normal with C1's but it did result in both patients desatting. Another event happened when vent was turned on, immediately received an alarm for "oxygen supply failed ". I switched the quick connect and that seemed to fix the issue. I took that same quick connect to the department to trial it on another unit but did not receive the same issue.
Patient problems: Low Oxygen Saturation
MAUDE report 50182648 ↗MalfunctionJul 1, 2026Improper Flow or Infusion
HAMILTON MEDIAL AG RECEIVED THE FOLLOWING COMPLAINT DESCRIPTION (SUMMARY): IT WAS REPORTED THAT THE HAMILTON-C1 VENTILATOR WAS ¿NOT GIVING FULL BREATHS¿ DURING PATIENT VENTILATION, ONLY 50¿60 ML AIR DELIVERED, NO LEAK FOUND, AND "HIGH RESPIRATORY RATE" ALARM WAS TRIGGERED. NO PATIENT, USER OR THIRD PARTY HARM NOR MEDICAL INTERVENTION WAS REPORTED. THE INVESTIGATION TO VERIFY THE ALLEGATION IS STILL IN PROGRESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25676704 ↗MalfunctionMay 21, 2026Application Program Problem
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: IT WAS REPORTED THAT THE DEVICE SHUTDOWN DURING VENTILATION AND ALARMED WITH TF 385002 (SAFETY FAILURE DETECTED), TF 485001 (AMBIENT FAILURE DETECTED), TF 443001 (WATCHDOG FAILED LLS). NO HARM TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25240038 ↗MalfunctionMay 5, 2026No Flow
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: THE DEVICE WAS REPORTED TO SHUT OFF DURING VENTILATION. PRIOR TO THE SHUTDOWN, TECHNICAL FAULTS TF 385002, 346041, 346054, AND 346040 WERE RECORDED. VENTILATION WAS CONTINUED USING AN ALTERNATIVE HAMILTON-C1. NO HARM TO THE PATIENT WAS REPORTED. THE LOGFILES WERE NOT RECEIVED. CURRENTLY, THE EVENT IS UNDER INVESTIGATION TO VERIFY THE REPORTED ALLEGATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25078882 ↗MalfunctionApr 2, 2026No Flow
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "NO FLOW WAS DELIVERED DURING HIGH FLOW VENTILATION." NO ALARMS WERE RAISED AND DEVICE "ACTED AS IF RUNNING NORMALLY." PATIENT WAS SWITCHED TO ANOTHER VENTILATOR. NO HARM TO THE PATIENT WAS REPORTED. CURRENTLY, THE INCIDENT IS UNDER INVESTIGATION TO VERIFY THE REPORTED ALLEGATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24762170 ↗InjuryApr 1, 2026Insufficient Device Problem Information
DEVICE GOES TO STANDBY MODE WHILE IT WAS CONNECTED TO PATIENT.
Patient problems: Insufficient Information
MAUDE report 24760661 ↗MalfunctionApr 1, 2026No Flow
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "NO DELIVERY OF GASES DURING HIGH FLOW O2 THERAPY. MEASURED GAS DELIVERY DISPLAYED ON SCREEN READS BETWEEN 0.0 LPM AND 0.3LPM. DEVICE DOES NOT ALARM OR INDICATE THAT THERE IS ISSUE WITH GAS DELIVERY." - NO HEALTH CONSEQUENCES OR IMPACT - NO INTERVENTION NECESSARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24755176 ↗InjuryMar 30, 2026No Flow
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: IT WAS REPORTED BY THE USER OF THE DEVICE THAT THE "DEVICE WENT INTO STANDBY MODE WITHOUT INTERACTION FROM STAFF." IT WAS FURTHER REPORTED THAT A CODE BLUE WAS INITIATED FOR THE PATIENT AND THAT THE PATIENT WAS TRANSFERRED TO THE EMERGENCY ROOM. INVESTIGATION IS ONGOING TO VERIFY THE ALLEGATIONS MADE IN THIS COMPLAINT (AS STANDBY MODUS IS NORMALLY THE RESULT OF 2 HUMAN INTERACTIONS WITH THE DEVICE). ADDITIONALLY, THE CASE WAS REPORTED BY THE USER TO THE COMPETENT AUTHORITY (FDA). THE REFERENCE NUMBER PROVIDED IS (B)(4).
Patient problems: Respiratory Arrest
MAUDE report 24725382 ↗MalfunctionFeb 25, 2026Failure to Run on Battery
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: QUOTATION FROM THE REPORTER: "AFTER INSTALLING NEW BATTERY S/N (B)(6) AND CHARGING AT LEAST 50%, VENT WAS ATTEMPTED TO BE USED ON A PATIENT. PRIOR TO THAT IT WAS ON BATTERY POWER AND PRODUCED TF CODES 446029, 431001, AND 485001.". IT WAS ADDITIONALLY MENTIONED THAT A DIFFERENT VENTILATOR HAD TO BE OBTAINED. NO HEATH CONSEQUENCE OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24439277 ↗MalfunctionFeb 19, 2026Obstruction of Flow
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: THE DEVICE ALARMED WITH EXHALATION OBSTRUCTED DURING VENTILATION. PATIENT WAS TRANSFERRED TO A DIFFERENT HAMILTON-C1. NO HARM TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24386996 ↗MalfunctionFeb 13, 2026Imprecision; Decrease in Pressure
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "LOSS OF PEEP CODED APPEARED. I CHECKED ALL CONNECTIONS, EVERYTHING WAS OKAY THE MASK HAD MINIMAL LEAK AT 7%. THEN MULTIPLE ERROR CODES APPEARED REALLY FAST AND THE UNIT SWITCHED TO CMV MODE ON ITS OWN AND WOULD NOT SWITCH BACK TO NIV ST" NO HEALTH CONSEQUENCES OR IMPACT. HOWEVER, BECAUSE THE RESPIRATORY TECHNICIAN WAS UNABLE TO CHANGE THE MODE, THEY REMOVED THIS UNIT FROM SERVICE. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24330223 ↗DeathFeb 6, 2026Improper Flow or Infusion
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "I AM WRITING THIS LETTER IN MY CAPACITY AS THE TREATING CLINICIAN TO FORMALLY SEEK A TECHNICAL EXPLANATION AND DEVICE EVALUATION REGARDING AN ADVERSE NEONATAL OUTCOME OBSERVED WHILE USING THE HAMILTON C1 NEO VENTILATOR IN OUR NICU. PATIENT & CLINICAL CONTEXT. PATIENT: MALE NEONATE, 2 DAYS OLD. CLINICAL SETTING: NICU. VENTILATOR USED: HAMILTON C1 NEO (NEONATAL MODE). INDICATION: RESPIRATORY DISTRESS WITH METABOLIC ACIDOSIS AND HYPOXEMIA. OBJECTIVE ABG EVIDENCE: DESPITE BEING ON INVASIVE MECHANICAL VENTILATION, SERIAL ABG REPORTS REVEALED PROGRESSIVE HYPOXEMIA AND WORSENING METABOLIC ACIDOSIS, AS DETAILED BELOW: EVENING ABG ((B)(6) 2025, 06:36 PM): PH: 7.202, PAO2: 24.8 MMHG, O2 SATURATION: 34.3%, HCO3¿: 13.6 MMOL/L, BASE EXCESS: ¿14.4 MEQ/L, ANION GAP: 26.5 MMOL/L. EARLIER ABG (SAME DAY): PH: 7.310, PAO2: 64.6 MMHG, HCO3¿: 15.8 MMOL/L, BASE EXCESS: ¿10.5 MMOL/L, O2 SATURATION: 91.1%. THESE REPORTS CLEARLY INDICATE PERSISTENT AND WORSENING HYPOXEMIA WHILE ON VENTILATOR SUPPORT, DESPITE STANDARD ESCALATION OF RESPIRATORY MANAGEMENT. CRITICAL CLINICAL OBSERVATION: A NOTABLE AND CONCERNING OBSERVATION WAS THAT: WHEN THE NEON
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information
MAUDE report 24275984 ↗MalfunctionFeb 4, 2026Application Program Problem
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "AS PER THE END USER, VENTILATOR HAS SHUT DOWN ITSELF WITH CONTINUOUS AUDIBLE ALARMS. AFTER THE END USER TURNED ON THE DEVICE, ALARMS STOPPED. THERE WERE TECHNICAL FAULTS. TF 346040, TF 346054, TF 346041, TF385002." IT WAS MARKED IN THE COMPLAINT THAT NO PATIENT WAS INVOLVED. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257594 ↗MalfunctionFeb 4, 2026No Flow
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: NO FLOW DELIVERED NO ALARMS RAISED DEVICE ACTED AS IF RUNNING NORMALLY. NO PATIENT INVOLVEMENT. NO INTERVENTION NECESSARY. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE, THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24255958 ↗MalfunctionFeb 3, 2026Failure to Deliver
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: WHEN THIS C1 VENTILATOR WAS BEING USED ON THE PATIENT IN PCV+ MODE, AN ALARM SUDDENLY SOUNDED, AND THE SCREEN DISPLAYED TE246015, 485001, 433001, AND 446029, AFTER WHICH THE VENTILATION OPERATION STOPPED. VENTILATOR HAD TO BE REPLACED. HOWEVER, THERE WAS NO HARM TO THE PATIENT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE, THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24243707 ↗MalfunctionJan 26, 2026Incorrect, Inadequate or Imprecise Result or Readings
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "CUSTOMER CALLED ABOUT THEIR HAMILTON C1 SN (B)(6) WHICH WAS ON A PATIENT. VENT HAD LOW MINUTE VOLUME AND LOW VOLUME ALARMS. VENT WAS AUTOCYCLING AND ALSO NOT DISPLAYING RESP. RATES OR VOLUME." VENTILATOR WAS EXCHANGED. NO HEALTH CONSEQUENCES OR IMPACT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE, THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24176004 ↗MalfunctionJan 21, 2026Failure to Deliver
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION:" WHILE USING THIS C1 ON A PATIENT, AN ALARM SUDDENLY SOUNDED, DISPLAYING ¿BLOWER FAILURE,¿ AND THE VENTILATOR SWITCHED TO AMBIENT MODE. THEREFORE, THE HOSPITAL STAFF IMMEDIATELY REPLACED THE VENTILATOR." NO HEALTH CONSEQUENCES OR IMPACT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24127968 ↗MalfunctionJan 14, 2026Defective Component
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "ERRROR 431002 BLOWER FAUILTY FOUND AT THE TIME OF CALIBATION PROCEES, SELF TEST FAILED". NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24065347 ↗MalfunctionJan 13, 2026Unexpected Mode Switch
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "WHILE USING THIS C1 ON A PATIENT, AN ALARM SUDDENLY SOUNDED AND ¿SAFETY VENTILATION¿ WAS DISPLAYED, SO THE HOSPITAL STAFF REPLACED THE VENTILATOR." THERE WERE NO HEALTH CONSEQUENCES OR IMPACT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE, THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24050693 ↗MalfunctionJan 6, 2026Unexpected Mode Switch
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "THE INCIDENT HAPPENED ON THE 16TH, HERE IS THE STATEMENT FROM THE RT. PLACED PT ON C1 (9074) FOR BIPAP, INITIAL EVERYTHING WAS OKAY, THEN IT SHOWED A LOSS OF PEEP ERROR, SO I CHECKED ALL CONNECTIONS, EVERYTHING WAS SECURE, THERE WAS MINIMAL LEAK ON THE MASK AT 7%, THEN THE C1 SHOWED DIFFERENT ERROR CODES AND SWITCHED INTO CMV ON ITS OWN, I TRIED TO SWITCH IT BACK TO NIV-ST BUT IT WOULD GO RIGHT BACK INTO CMV ON ITS OWN, I SWITCHED OUT MACHINE ON PATIENT FOR A NEW ONE." NO HEALTH CONSEQUENCES OR IMPACT - REPLACED HAMILTON-C1. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23972324 ↗MalfunctionNov 26, 2025Image Display Error/Artifact; Image Display Error/Artifact
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "CALL RECEIVED FROM CUSTOMER. DURING VENTILATION PARAMATERS ARE NOT SHOWING PROPERLY ON SCREEN. REPLACED VENTILATOR. LOG FILES UNABLE TO TAKE ON SUCH CONDITION. NO HEALTH CONSEQUENCES OR IMPACT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET. SO FAR NO RELATION TO THE DEVICE HAS BEEN ESTABLISHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23654247 ↗MalfunctionNov 26, 2025Pressure Problem; Pressure Problem
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "LOSS OF PEEP, HIGH BREATH RATE. MOVED PATIENT TO ANOTHER VENTILATOR. VENTILATOR SERVICED BY ONSITE BIOMED ON 10-02-2025. ACCOUNT MANAGER COMPLETED FUNCTIONAL TESTING 11/17 WITH NO ISSUES." NO HEALTH CONSEQUENCES OR IMPACT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE, THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET. SO FAR, NO RELATION TO THE DEVICE HAS BEEN ESTABLISHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23654165 ↗MalfunctionNov 4, 2025Overheating of Device; Overheating of Device
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "DRIVER BOARD EMITTED SMOKE THROUGH THE BATTERY PORTS WHEN POWERED ON. CUSTOMER WANTS FULL EVAL." NO PATIENT INVOLVEMENT. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23462040 ↗
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page