Skip to content

Bipolar Heads 25060.2850Recalls & reported problems

Medacta International SA · Ø28x50 · Class 2 · Implant

All models of Bipolar Heads (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630030843716), received between Sep 2, 2019 and Jun 6, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated3
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Pain2
  • Implant Pain1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 6, 2025Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem

    THE PATIENT HAD A LINK (COMPETITOR) PROXIMAL FEMORAL REPLACEMENT (STEM AND HEAD) WITH A MEDACTA BIPOLAR SHELL. ON THE (B)(6) 2025, AT ABOUT 1 MONTH FROM THE PRIMARY (ON (B)(6) 2025), THE PATIENT DISLOCATED BIPOLAR HEAD FROM THE NATIVE ACETABULUM. THE SURGEON PERFORMED AN OPEN SURGICAL PROCEDURE TO RELOCATE THE PROSTHESIS AND DURING THE CASE WENT FROM A 49 TO A 50MM BIPOLAR SHELL FOR ADDED STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY. COMPLAINT (B)(4) WAS FILLED. ON (B)(6) 2025, THE PATIENT HAD PAIN AFTER HIP LUXATION BECAUSE OF THE LACK OF ABDUCTOR TENSION. BECAUSE THE PATIENT HAD A HEMIARTHROPLASTY, THE SURGEON REMOVED ACETABULAR OSTEOPHYTES, OPENED THE ACETABULUM MEDIALLY, AND IMPLANTED A SMALLER BIPOLAR SHELL. HE ALSO CHANGED THE ANTEVERSION OF PROXIMAL FEMORAL REPLACEMENT TO AID WITH HIP STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Implant Pain

    MAUDE report 22156484 ↗
  • InjuryJan 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    REVISION SURGERY 3 DAYS POST PRIMARY (PATIENT WAS STILL AT THE HOSPITAL) DUE TO DISSOCIATION OF THE INOX HEAD FROM THE BIPOLAR HEAD. IT WAS ALSO REPORTED THAT THE AMISTEM-C STEM WAS FOUND LOOSE. ALL SYSTEM REVISED WITH SAME REFERENCES. NO TRAUMATIC EVENT WAS REPORTED AND NO COMPLICATIONS OCCURRED DURING THE PRIMARY SURGERY; THERE WERE NO MEDACTA EMPLOYEES PRESENT AT EITHER THE PRIMARY OR THE REVISION SURGERY.

    Patient problems: Pain

    MAUDE report 21167384 ↗
  • InjuryAug 26, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AT 23 DAYS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE BIPOLAR CUP AND FEMORAL HEAD FROM THE ACETABULUM. THE SURGEON REVISED THE BIPOLAR CUP AND FEMORAL HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 15300397 ↗
  • InjurySep 2, 2019Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REMOVED ALL IMPLANTS AND PUT IN NEW HARDWARE 3 WEEKS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 8953779 ↗

Other Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented devices

All Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented devices →

Sources and updates