Skip to content

M.U.S.T. MINI Polyaxial Screw 03.75.002Recalls & reported problems

Medacta International SA · Polyaxial Screw 3.5 x 14 Full thread + nut · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630030863165), received between Nov 25, 2020 and May 15, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Difficult to Insert1
  • Inadequacy of Device Shape and/or Size1

Patient problems

  • Bone Fracture(s)2
  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    C2-C7 FIXATION SURGERY. PATIENT'S PEDICLE BONE CRACKED AT C6 RIGHT SIDE WHILE INSERTING THE 3.5X14 PEDICLE SCREW (2.7 DIAMETER TAPPING PERFORMED). STRONG BONE. C6 LEVEL WAS SKIPPED AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 22025605 ↗
  • InjuryOct 17, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DURING THE SPINE SURGERY, A BONE FRACTURE WAS OBSERVED WHILE INSERTING THE MUST MINI POLYAXIAL SCREW 3.5 X 14. ANOTHER SCREW WAS USED, ALSO CHANGING THE SCREW TRAJECTORY. THE SURGERY HAS BEEN COMPLETED SUCCESSFULLY. 10 MINUTES DELAY. TOTAL SURGERY TIME 4 HOURS. THE BONE SEEMD TO BE HARD.

    Patient problems: Bone Fracture(s)

    MAUDE report 20469756 ↗
  • MalfunctionMay 24, 2022Difficult to Insert; Difficult to Insert

    DURING THE PRIMARY SPINE SURGERY, DURING FINAL IMPLANTATION, WHEN THE SURGEON WAS TORQUING OUT THE RODS, 3 OF THE SET SCREWS WERE NOT ABLE TO BE TORQUED DOWN. THE SURGEON HAD TO REMOVE THE SCREWS AND IMPLANT NEW SCREWS. THERE WAS A 30-MINUTE DELAY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. A CONSIGNMENT SET WAS UTILIZED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14485015 ↗
  • InjuryNov 25, 2020Inadequacy of Device Shape and/or Size

    SPINAL FIXATION WAS PERFORMED AT C2-T2. OPEN SURGERY. THE BONE QUALITY SEEMED TO BE LOW. WHEN THE SURGEON ENGAGED THE CERVICAL SCREW WITH THE ROD AT LEFT SIDE OF C2, THE CERVICAL SCREW WAS PULLED OUT FROM THE PATIENT BODY BECAUSE THE CERVICAL SCREW WAS NOT ABLE TO WITHSTAND THE LOAD. THE SURGEON DECIDED TO DO NOT IMPLANT THE SCREW IN C2 LEFT. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 30 MINUTES. TOTAL SURGERY TIME WAS 280MINS.

    Patient problems: Failure of Implant

    MAUDE report 10905039 ↗

Other Orthosis, Cervical Pedicle Screw Spinal Fixation devices

All Orthosis, Cervical Pedicle Screw Spinal Fixation devices →