MectaLIF Anterior System 03.30.132Recalls & reported problems
Medacta International SA · Enhanced Screw Diam.5x30 (Double Packaging) · Class 2 · Implant
All models of MectaLIF Anterior System (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630030875427), received between May 25, 2018 and Jul 5, 2019 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component1
- Device Slipped1
- Fracture1
Patient problems
- No Known Impact Or Consequence To Patient4
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 5, 2019Fracture; Fracture
SCREW BREAKAGE AFTER ABOUT 1 YEAR FROM PRIMARY SURGERY. NO REVISION SURGERY HAS BEEN PERFORMED FOR THE MOMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8764307 ↗MalfunctionMay 10, 2019Device Slipped
THE SURGEON DISCOVERED THAT ONE OF THE SCREW BACKED OUT. THE SURGEON DOESN'T REVISE THE SCREW BUT HE IS MONITORING THIS PATIENT TO UNDERSTAND IF THE SCREW IS CONTINUING TO BACK-OUT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8599631 ↗MalfunctionNov 19, 2018Break; Detachment of Device or Device Component
FROM THE FINAL PER-OP X-RAYS THE SURGEON DETECTED THAT THE PLATE WAS PROUD BY 1MM FROM THE ANTERIOR PORTION OF THE VERTEBRAL BODY. HE PROCEEDED TO USE THE FIXED STRAIGHT SCREW DRIVER TO BRING THE PLATE DOWN AND BY OVERNIGHTING THE SCREW, THE HEAD SHEARED OFF. THERE WERE NO FRAGMENTS, IT WAS A CLEAN BREAK. THE SCREW SHANK WAS LEFT IN PLACE, BURIED/COUNTERSUNK IN THE CAGE/BONE. THE BROKEN SCREW HEAD WAS REMOVED. NO DEFORMATION OR BREAKAGE OF THE INSTRUMENTS OCCURRED. THERE WERE A FIVE MINUTES OF DELAY IN THE CASE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8084558 ↗MalfunctionSep 20, 2018Break; Break
THE PATIENT COMPLAINED OF BACK PAIN ON (B)(6) 2018 AND WAS SENT FOR X-RAYS (L5/S1 ALIF PROCEDURE). THIS SHOWED BOTH S1 SCREWS HAD BROKEN. THE SURGEON HAS DECIDED NOT TO REVISE THIS AND ALSO TO CHANGE HIS TECHNIQUE TO INCLUDE DRILLING AND ALSO MORE X-RAYS DURING PROCEDURE TO ENSURE NO GAP. HE IS SATISFIED THAT THE CAGE IS FUSING WELL AND THAT THE PAIN IS NOT COMING FROM THIS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7893365 ↗InjuryMay 25, 2018Break
THE PATIENT HAD A PRIMARY ALIF AT L5-S1. THE PATIENT CAME IN COMPLAINING OF PAIN. THE SURGEON DETERMINED THAT THE SCREW BROKE. THERE IS CURRENTLY NO PLAN TO REVISE THE PATIENT.
Patient problems: Pain
MAUDE report 7544747 ↗
Other Intervertebral Fusion Device With Integrated Fixation, Lumbar devices
- PIVOX™ Oblique Lateral Spinal System 2161001Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, lumbar182 reports
- IdentiTi 232-10-85-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall8 reports
- IdentiTi 232-10-85-250Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall5 reports
- IdentiTi 132-10-85-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall3 reports
- MectaLIF Anterior 03.30.301Medacta International SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall2 reports
- IdentiTi 232-10-85-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall2 reports
- IdentiTi 232-11-90-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall1 report
- MectaLIF Anterior 03.30.302Medacta International SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page