MectaLIF Anterior 03.30.300Recalls & reported problems
Medacta International SA · MectaLIF Ant - Lag Plate Flush H10 mm · Class 2 · Implant
All models of MectaLIF Anterior (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630345732286), received between Sep 9, 2024 and Sep 9, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 8, 2023Class IICompleted
There is a potential for breakage of the small screw used to affix the anti-backout plate.
Z-1400-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2024Fracture; Fracture
THE ANTI-BACKOUT SCREW CHIPPED OFF IN THE IMPLANT AND THE SURGEON WAS NOT ABLE TO INSERT IT. HE LEFT THE BROKEN THREADED PORTION OF THE SCREW IN THE IMPLANT AND THEN USED A NON-MEDACTA ANTERIOR PLATE AS A MEANS TO PREVENT SCREW BACK OUT. THE SURGEON CONFIRMED THAT THE SCREW SHEARED OFF AS HE WAS ATTEMPTING TO TIGHTEN IT USING THE STANDARD TORQUE DRIVER. HE DID NOT USE A MALLET OR ANY OTHER INSTRUMENT. THERE WAS A 5-MINUTE DELAY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20167667 ↗
Other Intervertebral Fusion Device With Integrated Fixation, Lumbar devices
- PIVOX™ Oblique Lateral Spinal System 2161001Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, lumbar182 reports
- IdentiTi 232-10-85-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall8 reports
- IdentiTi 232-10-85-250Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall5 reports
- IdentiTi 132-10-85-300Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall3 reports
- MectaLIF Anterior 03.30.301Medacta International SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall2 reports
- IdentiTi 232-10-85-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall2 reports
- IdentiTi 232-11-90-350Alphatec Spine, Inc. · Intervertebral fusion device with integrated fixation, lumbar1 recall1 report
- MectaLIF Anterior 03.30.302Medacta International SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.