Knee Instruments 11.00004Recalls & reported problems
Medacta International SA · Conventional instrument set · Class 1
All models of Knee Instruments (33)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030877377), received between Jan 16, 2019 and Jan 18, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Device Fell1
- Material Fragmentation1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 18, 2019Break; Material Fragmentation
DURING THE PRIMARY KNEE SURGERY, AND WHILE THE SURGEON WAS REMOVING THE IM ROD AND POSITIONING THE DISTAL FEMORAL CUTTING GUIDE, THE PLASTIC SLEEVE OF THE INSTRUMENT BROKE. THE SURGEON HAD TO USE AN EXCESSIVE AMOUNT OF FORCE TO REMOVE THE IM ROD AND NO DEFECTS WERE OBSERVED IN THE IM ROD PRIOR TO INSERTION. FRAGMENTS FROM THE PLASTIC SLEEVE WERE LEFT IN THE PATIENT'S INTER-MEDULLARY CANAL AND THE SURGEON WAS UNABLE TO RECOVER THE FRAGMENTS. ONE FRAGMENT WAS 5CM IN LENGTH. THERE WAS A 35 MINUTE DELAY IN THE CASE AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Foreign Body In Patient
MAUDE report 8260589 ↗MalfunctionJan 16, 2019Break; Device Fell
DURING THE PRIMARY KNEE SURGERY, THE PLASTIC SLEEVE ON THE IM ROD BROKE WHILE THE SURGEON WAS PREPARING TO PERFORM THE DISTAL CUT ON THE FEMUR. A PIECE OF PLASTIC FELL INTO THE PATIENT'S FEMORAL CANAL AND THE SURGEON WAS UNABLE TO RETRIEVE IT. THERE WAS A TEN MINUTE DELAY IN THE CASE AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8253139 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page