APTUS A-5750.14/1SRecalls & reported problems
Medartis AG · 2.5 TriLock Screw 14mm, HD7 · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630037882060), received between Apr 1, 2026 and Apr 1, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Insufficient Device Problem Information1
Patient problems
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 1, 2026Break; Insufficient Device Problem Information
AT (B)(6) HOSPITAL IN (B)(6), A SURGEON PERFORMING WRIST SURGERY UTILIZED A 2.5 TRILOCK ULNA SHORTENING PLATE IN COMBINATION WITH A 2.5 TRILOCK SCREW. DURING INSERTION OF THE 2.5 TRILOCK SCREW INTO THE SECOND DISTAL HOLE OF THE PLATE AND SUBSEQUENT LOCKING WITH A DRIVER, THE SCREW HEAD SEPARATED. THE BROKEN SCREW REMAINED IN SITU, AND THIS INCIDENT WILL BE REPORTED TO THE PMDA. INTRAOPERATIVELY, THE SURGEON ASSESSED THAT REMOVAL OF THE BROKEN SCREW DURING THE SAME PROCEDURE WOULD BE CHALLENGING. CONSEQUENTLY, THE DECISION WAS MADE TO LEAVE THE FRAGMENT IN PLACE, WITH PLANS TO EMPLOY A RESCUE KIT OR A COMPARABLE SOLUTION FOR EXTRACTION AT THE TIME OF SCREW REMOVAL. THE SURGEON REPORTED THAT THE PATIENT'S BONE QUALITY WAS HARD. THE SCREW WAS PRE-DRILLED WITH A SUITABLE TWIST DRILL, AND BASED ON THE AVAILABLE INFORMATION, EXCESSIVE FORCE WAS NOT APPLIED DURING INSERTION. NO ADDITIONAL X-RAYS ARE AVAILABLE FOR ANALYSIS. DESPITE THIS INTRAOPERATIVE COMPLICATION, THE SURGEON CONFIRMED THAT THE INTENDED SURGICAL OUTCOME WAS ACHIEVED.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 24757002 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page