ACDF Stand Alone Cervical 03.18.110Recalls & reported problems
Medacta International SA · Mecta-C SA TiPEEK Cage 16x14x6x7° · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630040725644), received between Apr 25, 2025 and Apr 25, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult or Delayed Positioning1
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 25, 2025Difficult or Delayed Positioning; Loose or Intermittent Connection; Difficult or Delayed Positioning; Loose or Intermittent Connection
DURING INSERTION OF THE LOCKING SCREW THE FLUSH LOCK PLATE CAME LOOSE FROM THE CAGE. THE SURGEON RECOUPLED THE PLATE AND THE CAGE AND TRIED AGAIN TO INSERT THE SCREW BUT THE PLATE CAME LOOSE AGAIN. NEW CAGE AND PLATE WERE USED TO COMPLETE THE SURGERY. THE SAME ISSUE OCCURRED AGAIN WITH A SECOND CAGE ON A DIFFERENT LEVEL: THE PLATE CAME LOOSE, THE SURGEON RECOUPLED IT TO THE CAGE AND IN THIS CASE WAS ABLE TO COMPLETE SCREW INSERION WITHOUT ISSUES. PATIENT WAS YOUNG WITH GOOD BONE QUALITY, DISTRACTION WAS PERFORMED WITH CASPAR DISTRACTOR. IT WAS REPORTED THAT THE SURGEON DID NOT TRY TO REPOSITION THE CAGES ONCE IN SITU AND BEFORE SCREWING. 20 TO 30 MINUTES OF DELAY REPORTED, TOTAL SURGERY TIME AROUND 2H.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21907807 ↗
Other Intervertebral Fusion Device With Integrated Fixation, Cervical devices
- NA 03.617.902Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical103 reports
- NA 03.617.900Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical77 reports
- Divergence® Anterior Cervical Fusion System G6626525Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical75 reports
- Divergence® Anterior Cervical Fusion System G6626526Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical49 reports
- NA 03.617.905Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical29 reports
- NA 03.617.904Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical27 reports
- Divergence® Anterior Cervical Fusion System G6626746Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical21 reports
- NA 03.647.972Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.