HIP Instruments 01.15.10.0133Recalls & reported problems
Medacta International SA · Straight AMIS Broach Handle · Class 1
All models of Hip Instruments (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630040735254), received between Sep 5, 2019 and Sep 5, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Device Fell1
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2019Device Fell
DURING THE FEMORAL RASPING, WHEN THE SURGEON PULL OUT THE STRAIGHT AMIS BROACH HANDLE WITH A HAMMER, THE REVER DISENGAGED AND FELL IN THE SURGICAL FIELD. THE SURGEON REMOVED THE BROACH FROM THE MEDULLARY CAVITY WITH ANOTHER INSTRUMENT. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 10 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8964797 ↗InjurySep 5, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
WHEN THE SURGEON TRIED TO REMOVE THE BROACH FROM THE MEDULLARY CAVITY WITH THE STRAIGHT AMIS BROACH HANDLE, THE LEVER DISENGAGED FROM THE STRAIGHT AMIS BROACH HANDLE. SEVERAL PARTS DISASSEMBLED AND SOME WERE FELL INTO THE SURGICAL FIELD. ALL OF THEM WERE ABLE TO RETRIEVE. THE SURGEON REMOVED THE BROACH FROM THE MEDULLARY CAVITY WITH ANOTHER INSTRUMENT. DUE TO THIS EVENT, THE SURGERY WAS PROLONGED ABOUT 10 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8964687 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page