HIP Instruments 01.22.10.0117Recalls & reported problems

Medacta International SA · Short Screwdriver 5mm · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630040739030), received between Jul 18, 2025 and Jul 18, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 18, 2025Fracture; Fracture

    DURING FINAL TIGHTENING OF THE DEFINITIVE SCREW THE TIP OF THE SHORT SCREWDRIVER 5MM BROKE INSIDE THE SCREW; IT COULD NOT BE RECOVERED AND WAS LEFT IN THE PATIENT. THE SCREW HAS BEEN REPORTED TO BE COMPLETELY TIGHTENED AND THE CONSTRUCT SEEMED STABLE. TORQUE LIMITER WAS USED AND WAS REPORTED TO HAVE REACHED THE TARGET TORQUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22541065 ↗