Skip to content

HIP Instruments 01.22.10.0151Recalls & reported problems

Medacta International SA · Proximal Body Handle · Class 1

All models of Hip Instruments (48)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630040739252), received between Mar 28, 2024 and Jun 5, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2025Material Twisted/Bent; Material Twisted/Bent

    AFTER PROXIMAL BODY INSERTION, THE SURGEON WASN'T HAPPY WITH THE STEM VERSION AND WANTED TO CHANGE IT. THE PROXIMAL BODY HANDLE WAS SCREWED ON THE NECK AND THE SMALL ROD PLACED INSIDE WITH THE T-HANDLE. BUT IT WAS IMPOSSIBLE TO REMOVE THE PROXIMAL BODY BECAUSE THE HANDLE WAS STARTING TO BEND. THE SURGEON CHANGED THE CUP TO A DUAL MOBILITY TO COMPENSATE FOR THE VERSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22149349 ↗
  • MalfunctionMar 28, 2024Material Deformation; Material Deformation

    DURING A REVISION SURGERY (DUE TO THE DISLOCATION OF THE BIPOLAR HEAD FROM THE ACETABULAR), AFTER COMPLETING THE ACETABULAR WORK, THE SURGEON ATTEMPTED TO DISASSEMBLE THE PROXIMAL BODY FROM THE DISTAL STEM, BECAUSE HE WANTED TO REVISE ALSO THE PROXIMAL STEM TO GIVE MORE STABILITY TO THE ARTICULATION. THE PROXIMAL BODY COULD NOT BE DISASSEMBLED AND THE DISASSEMBLY ROD AND THE PROXIMAL BODY HANDLE WERE DEFORMED. FOLLOWING THIS ATTEMPT, THEY USED THE BACKUP HANDLE AND ANOTHER DISASSEMBLY ROD, AND AGAIN THE COMPONENTS COULDN'T BE UNCOUPLED. AT THIS POINT, THEY ATTEMPTED TO BACK OFF THE BODY WITH THE STEM REMOVAL INSTRUMENTS BUT IT WOULD NOT RELEASE. THE SURGEON THEN DECIDED TO LEAVE THE M-VIZION STEM. ONLY 5 MINUTES DELAY REPORTED. THE SURGEON WAS SATISFIED WITH THE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18993010 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates