Hip Instruments 01.10.10.131Recalls & reported problems
Medacta International SA · Offset-30° broach handle (left) · Class 1
All models of Hip Instruments (48)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07630040742122), received between Feb 28, 2019 and Apr 7, 2021 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break6
- Material Fragmentation2
- Detachment of Device or Device Component1
- Device Fell1
Patient problems
- No Known Impact Or Consequence To Patient8
- Insufficient Information1
- No Code Available1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 7, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
DURING THE SAME SURGERY THE BROACHES HANDLES DISASSEMBLED WHILE THE SURGEON WAS SWITCHING FROM A BROACH TO THE NEXT ONE. THE SURGERY WAS COMPLETED SUCCESSFULLY USING EXTRA STRAIGHT BROACH HANDLES, BUT THERE WERE 40 MINUTES OF DELAY DUE TO THIS ISSUE.
Patient problems: Insufficient Information
MAUDE report 11626168 ↗MalfunctionAug 30, 2019Material Fragmentation; Material Fragmentation
THE BALL AND THE SPRING OF THE LEVER FROM THE OFFSET BROACH HANDLE WAS FALLING IN THE PATIENT DURING THE SURGERY. THE PART ARE REMOVED FROM THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8949450 ↗MalfunctionAug 22, 2019Break; Break
BALL BEARING MECHANISM BREAKAGE DURING SURGERY. THE BALL DID NOT FALL INTO THE PATIENT. A STRAIGHT BROACH HANDLE WAS USED TO COMPLETE THE CASE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8921003 ↗MalfunctionJun 14, 2019Material Fragmentation; Material Fragmentation
THE LOCKING BALL OF THE INSTRUMENT LEVER FELL OUT FROM THE HANDLE. NO PIECE FELL INTO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8699598 ↗MalfunctionApr 26, 2019Break; Break
DURING THE IMPACTION THE INSTRUMENT BALL SPRING FELL OUT THE PASSING HOLE AND THE SPRING WAS BROKEN. THE BALL DID NOT FALL IN THE PATIENT. THE SURGERY WAS COMPLETED SUCCESSFULLY USING A STRAIGHT BROACH HANDLE. THE SAME DAY THE SURGEON HAD A SECOND IDENTIC CASE WITH A SECOND BROACH HANDLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8555124 ↗MalfunctionMar 21, 2019Break; Break
DURING THE SURGERY THE LOCKING BALL FELL OUT OF THE HANDLE. THE BALL IS NOT FALLING IN THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8439896 ↗MalfunctionMar 18, 2019Break; Break
THE LOCKING BALL FELL OUT OF THE HANDLE. A STRAIGHT BROACH HANDLE (REF. 01.15.10.0131) WAS USED TO COMPLETE THE CASE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8428910 ↗MalfunctionMar 1, 2019Device Fell
DURING THE INSTRUMENT IMPACTION THE LOCKING BALL AND THE SPRING FALL OUT THE PASSING HOLE. THE BALL DIDN'T FALL IN THE PATIENT. THE SURGEON FINISHED THE IMPACTION WITH THE SAME HANDLE WHILE HE HOLDS THE LEVER WITH HIS HAND.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8381775 ↗MalfunctionFeb 28, 2019Break; Break
DURING THE IMPACTION THE BALL OF THE BROACH HANDLE FELL OUT FROM THE PASSING HOLE AND THE SPRING WAS BROKEN, NO DEBRIS FELL IN THE PATIENT. THERE WAS A DELAY OF MAX. FIVE MINUTES.
Patient problems: No Code Available
MAUDE report 8377334 ↗MalfunctionFeb 28, 2019Break; Break
DURING SURGERY THE LOCKING BALL AND THE SPRING FELL OUT OF THE HANDLE WHILE A NURSE WANTED TO REMOVE THE BROACH FROM THE HANDLE. THERE WAS A DELAY OF MAX. FIVE MINUTES. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8377333 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page