Hip Instruments 01.10.10.190Recalls & reported problems
Medacta International SA · Stem Repositioner · Class 1
All models of Hip Instruments (48)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630040746809), received between Feb 15, 2019 and Oct 11, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 11, 2024Material Integrity Problem; Material Integrity Problem
DURING HIP PRIMARY SURGERY, THE SURGEON WANTED TO REMOVE THE STD SIZE 4 STEM AS IT WAS IMPLANTED A WRONG SIZE RESULTING IN A TOO HIGH IMPLANT AND IMPLANT A SIZE 3. HOWEVER, THE FIXATION WAS GOOD AND THE SURGEON COULD NOT GET THE STEM OUT, USING THE STEM REPOSITIONER AND SCREWED STEM EXTRACTOR. AFTER REPEATED ATTEMPTS, THE STEM REPOSITIONER HAD WORN OUT AND THE STEM WAS LEFT IN PLACE. THE DEVICE WAS NOT DAMAGED BEFORE USING IT. CUP SIZE S AND STD SIZE 4 STEM HAS BEEN IMPLANTED, WITH THE PLANNED OFFSET. ONE HOUR DELAY OCCURRED TO RELEASE ALL LIGAMENTS TO INCREASE THE LEG'S LEGHT (TOTAL TIME SUGERY 2HR AND 15MIN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20426603 ↗MalfunctionFeb 15, 2019Break
DURING THE PRIMARY HIP SURGERY AND AFTER ALL COMPONENTS WERE IMPLANTED, THE SURGEON WAS NOT ABLE TO REDUCE THE PATIENT. THE SURGEON ATTEMPTED TO EXTRACT THE STEM WITH THE STEM REPOSITIONER BUT THE STEM WAS WELL FIXED AND HE WAS NOT ABLE TO EXPLANT THE STEM. DURING THE PROCESS OF ATTEMPTING TO EXTRACT THE STEM, THE STEM REPOSITIONER WAS STRUCK WITH A MALLET AND IT BROKE. AFTER FURTHER EXAMINATION, THE SURGEON DETERMINED THAT THE REASON HE WAS NOT ABLE TO REDUCE THE PATIENT WAS THAT THE STEM WAS CONTACTING THE ACETABULUM. THE SURGEON THEN REMOVED A SMALL PORTION OF BONE FROM THE ACETABULUM AND THE PATIENT WAS EASILY REDUCED. THERE WAS A 40-MINUTE DELAY IN THE CASE AND THE SURGERY WAS COMPLETED SUCCESSFULLY. NO FRAGMENTS FELL INTO THE PATIENT. ADDITIONAL INFORMATION: THE INSTRUMENT BROKE AFTER BEING STRUCK WITH MALLET: THE CIRCLED AREA IS THE BROKEN PART OF THE INSTRUMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8343137 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page