Hip Instruments 01.18.10.102Recalls & reported problems
Medacta International SA · Trial Broach AMIStem # 2 · Class 1
All models of Hip Instruments (48)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630040751780), received between Apr 5, 2019 and Feb 12, 2021 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture2
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Mechanical Problem1
Patient problems
- Bone Fracture(s)3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 12, 2021Fracture; Appropriate Device Problem Term/Code Not Available
THE SURGEON DISCOVERED THE FRACTURE AROUND THE FEMORAL STEM BY CT-IMAGE DURING FEMORAL BONE RASPING. THE SURGEON FIXED THE FRACTURE WITH THE NESPLON TAPE, AND INSERTED THE AMISTEM-P SHORT NECK STD SIZE 4. DAA SURGERY. AMIS LEG POSITIONER WAS USED. LOT NUMBER OF THE INSTRUMENT IS UNAVAILABLE.
Patient problems: Bone Fracture(s)
MAUDE report 11322060 ↗InjuryAug 28, 2019Fracture
AFTER FEMORAL RASPING, THE SURGEON TRIED TO REPOSITION AND FELT AN ABNORMAL FEELING. CHECKING THE CT IMAGE HAS BEEN DISCOVERED THE TRANSVERSE FRACTURE ABOUT 10MM FURTHER THAN THE DISTAL RASPING POINT. THE SURGEON FIXED THE FRACTURE AND CABLE. DUE TO THIS EVENT, THE SURGERY WAS PROLONGED ABOUT 2 HOURS. TOTAL SURGERY TIME WAS ABOUT 4 HOURS. THE RASP DID NOT SHOW ANY DEFORMATIONS OR SCRATCHES.
Patient problems: Bone Fracture(s)
MAUDE report 8939223 ↗InjuryApr 5, 2019Break; Mechanical Problem
DURING SURGERY THE FEMORAL BONE BROKE WHEN THE SURGEON TRIED TO ENGAGE THE HANDLE WITH THE RASP. FEMORAL BONE WAS FIXED WITH THE WIRE. THE SURGEON COMMENTED THAT IT WAS HARD TO ENGAGE THE HANDLE WITH THE RASP.
Patient problems: Bone Fracture(s)
MAUDE report 8486483 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page