Hip Instruments 01.20.10.100Recalls & reported problems

Medacta International SA · Head Reamer Ø 8 · Class 1

All models of Hip Instruments (48)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630040758543), received between Aug 30, 2023 and Aug 30, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Ejection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Ejection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2023Ejection Problem; Ejection Problem

    THE DISTAL REAMER DETACHED FROM THE LONG RIGID SHAFT FOR REAMERS AND REMAINED STUCK IN THE DISTAL FEMUR DURING AMIS-K FEMUR PREPARATION.THE SURGEON TURNING THE REAMER IN COUNTERCLOCKWISE DIRECTION (THE INSTRUMENT CUT IN CLOCKWISE DIRECTION). 30 MINUTES OF DELAY IN THE SURGERY TO RECOVER THE REAMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17648994 ↗