Skip to content

Knee Instruments 02.07.10.4048Recalls & reported problems

Medacta International SA · Safe guiding reamer Ø11 · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630040760119), received between Sep 27, 2024 and Sep 27, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 27, 2024Use of Device Problem; Use of Device Problem

    WHEN USING THE CROSSOVER INSTRUMENTATION, IT WAS NOTICED THAT THE REAMERS IN THE SLEEVES DO NOT HAVE PROPER GUIDANCE AS SOON AS THEY ARE INSERTED INTO THE SHAFT AND TEND TO DEVIATE. THIS CREATED PROBLEMS WHEN PLACING THE TRIALS AND THE FINAL IMPLANTS. THE SURGEON FIRST USED SAFE GUIDING REAMER Ø11MM, THEN SAFE GUIDING REAMER Ø13MM, ANF INALLY SAFE GUIDING REAMER Ø16MM. THE OPERATION TIME WAS EXTENDED BY ABOUT AN HOUR AND MORE ANESTHETIC HAD TO BE ADMINISTERED. ON THE TIBIA, THE TIBIA WAS INITIALLY NOT PLACED CLEANLY ON THE RESECTION BECAUSE THE SHAFT WAS DIVERTED. THE FEMUR WAS FLEXED AND HAD NO SECURE HOLD BY THE SHAFT. THIS HAD TO BE HELD IN PLACE FOR CEMENTATION, OTHERWISE IT WOULD NOT HAVE REMAINED IN THE CORRECT POSITION. NO DEFECTS CAN BE NOTED ON THE REAMERS. THE PATIENT WAS VERY YOUNG AND HAD HARD BONES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20320103 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page