Skip to content

HIP Instruments 01.39.10.0033Recalls & reported problems

Medacta International SA · Calcar Reamer Large Zimmer Hall · Class 1

All models of Hip Instruments (48)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630040763790), received between Jan 12, 2024 and Jun 27, 2024 — classified as 3 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Bone Fracture(s)3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    FEMUR FRACTURE CAUSED WHEN THE SURGEON WAS USING THE LARGE CALCAR REAMER AND THE BROACH. THE SURGEON CABLED THE FRACTURE COMPLETING SUCCESSFULLY THE CASE. THE STEM IMPLANTED WAS AN AMISTEM-P.

    Patient problems: Bone Fracture(s)

    MAUDE report 19621813 ↗
  • InjuryJun 21, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DURING CALCAR REAMING THE, CALCAR REAMER CAUGHT ON X-ACTA BROACH WHICH CAUSED IT TO TORQUE AND RESULTED IN A LONGITUDINAL PROXIMAL FEMUR FRACTURE. CABLES WERE USED TO REDUCE THE FRACTURE. THE SURGEON WAS USING AXIAL FORCE WITH THE CALCAR REAMER ON POWER. HE SUSPECTS THE FORCE MAY HAVE BEEN OFF AXIS AND THE TEETH OF THE CALCAR REAMER CAME INTO CONTACT WITH THE PROXIMAL PART OF THE BROACH. THE INSTRUMENTS WERE ALL IN GOOD CONDITION WITH NO DEFORMATION/ DAMAGE PRESENT.

    Patient problems: Bone Fracture(s)

    MAUDE report 19582673 ↗
  • InjuryJan 12, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DURING THE PRIMARY HIP SURGERY, A FEMORAL BONE FRACTURE OCCURRED WHILE REAMING THE CALCAR WITH THE CALCAR REAMER LARGE ZIMMER HALL. THE SURGERY WAS COMPLETED SUCCESSFULLY. NOTHING WAS DONE TO THE FRACTURE, AS IT WAS "CONTAINED" AND THE SURGEON DID NOT FEEL THAT A CABLE OR PLATE WAS NECESSARY.

    Patient problems: Bone Fracture(s)

    MAUDE report 18501924 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates