Spine Instruments 03.30.10.0013Recalls & reported problems
Medacta International SA · ALIF Tap - 5mm · Class 1
All models of Spine Instruments (28)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630040798396), received between Nov 19, 2018 and Apr 5, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture2
- Material Fragmentation1
Patient problems
- No Known Impact Or Consequence To Patient2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2019Material Fragmentation; Material Fragmentation
DURING THE VERTEBRA TAPPING, IN THE ANTERIOR SURGERY, THE TIP OF THE INSTRUMENT BROKE OFF. THE BROKEN FRAGMENT REMAINED IN SITU AND FOR THIS THE SURGEON WAS ABLE TO INSERT ONLY 3 BONE SCREWS. 9 DAYS AFTER PRIMARY SURGERY THE SURGEON PERFORMED A SECOND SURGERY TO INSERT THE POSTERIOR STABILIZATION IMPLANTS. THE SURGERY WAS COMPLETED ANYWAY SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8486482 ↗MalfunctionMar 12, 2019Fracture; Fracture
WHEN TAPPING THE PATHWAY FOR THE FIRST SCREW, THE TIP OF THE TAP BROKE OFF. ACCORDING TO THE SURGEON NO EXCESSIVE LATERAL FORCES WERE APPLIED. THE TIP WAS REMOVED WITH THE HELP OF A KOCHER FORCEPS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8412845 ↗MalfunctionNov 19, 2018Fracture; Fracture
WHEN USING THE TAP STRAIGHT IN THE USUAL MANNER TO PREPARE THE HOLE FOR THE SCREW, THE TIP OF THE TAP BROKE OFF. THE BROKEN PIECE REMAINED IN THE PATIENT BUT TO DATE THERE WAS NO NECESSITY TO DO ANOTHER SURGERY AND THERE WAS NO DELAY IN THE COURSE OF THE CURRENT SURGERY AS WELL.
Patient problems: Foreign Body In Patient
MAUDE report 8084541 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page