Skip to content

Knee Instruments 02.07.10.8651Recalls & reported problems

Medacta International SA · 3mm trial offset with locking system · Class 1

All models of Knee Instruments (33)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07630345713148), received between Aug 2, 2023 and Mar 31, 2026 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Fracture” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture9

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Bone Fracture(s)1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Fracture

    THE ANTIROTATIONAL INSERT OF THE TRIAL OFFSET 3MM BROKE AND REMAINED ATTACHED THE TIBIAL PUNCHER, WHILE THE BODY OF THE TRIAL OFFSET REMAINED STUCK IN THE TIBIAL CANAL. ABOUT 45 MINUTES WERE REQUIRED TO REMOVE THE BROKEN PART AND THE SURGEON HAD TO REMOVE PART OF THE TIBIAL BONE TO BE ABLE TO GRAB THE OFFSET. BROKEN PART RETRIEVED AND SURGERY COMPLETED SUCCESSFULLY. BONE QUALITY NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24738824 ↗
  • MalfunctionJan 29, 2026Fracture

    THE ANTI-ROTATIONAL INSERT OF THE TRIAL OFFSET 3MM BROKE DURING THE REMOVAL OF THE TRIAL ASSEMBLY. CLAMP USED TO REMOVE THE PART (5 MINUTES DELAY TO RECOVER THE PIECE). SURGERY COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24208585 ↗
  • MalfunctionJul 4, 2025Fracture; Fracture

    THE ANTIROTATIONAL INSERT OF THE TRIAL OFFSET 3MM BROKE. BROKEN PART RETRIEVED AND SURGERY COMPLETED SUCCESSFULLY. 1 HOUR DELAY OCCURRED TO REMOVE THE BROKEN PART. THE SURGEON HAD TO CUT THE DIAPHYSIS TO BE ABLE TO GRAB THE OFFSET AND PULL OUT THE OFFSET AND STEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22403097 ↗
  • MalfunctionFeb 20, 2025Fracture; Fracture

    THE ANTIROTATIONAL INSERT OF THE TRIAL OFFSET 3MM BROKE AND REMAINED INSIDE THE TRIAL FEMUR AND STEM. THE OPERATION WAS EXTENDED BY AROUND 30 MINUTES AND THE PATIENT HAD TO REMAIN UNDER ANESTHESIA LONGER. IN ADDITION, BONE HAD TO BE REMOVED FROM THE ANTERIOR FEMORAL BONE ABOVE THE OFFSET CONNECTOR IN ORDER TO BE ABLE TO GRASP THE SHAFT WITH FORCEPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21420549 ↗
  • MalfunctionNov 8, 2024Fracture; Fracture

    THE ANTIROTATIONAL INSERT OF THE TRIAL OFFSET 3MM BROKE AND REMAINED INSIDE THE TRIAL STEM. THE SURGEON DRILLED HOLES IN THE TRIAL OFFSET IN ORDER TO HOOK IT AND PULL IT OUT. SURGERY WAS COMPLETED WITH 1 HOUR OF DELAY DUE TO THE ISSUE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20639490 ↗
  • MalfunctionMay 6, 2024Fracture; Fracture

    THE ANTI-ROTATIONAL INSERT OF THE TRIAL OFFSET BROKE. THE OFFSET TRIAL AND THE STEM WERE BLOCKED IN THE INTRAMEDULLARY CANAL. THE BROKEN PART WAS REMOVED AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19248897 ↗
  • InjuryDec 4, 2023Fracture; Fracture

    THE ANTI-ROTATIONAL INSERT OF THE TRIAL OFFSET 3MM BROKE DURING THE REMOVAL OF THE TRIAL ASSEMBLY. TO REMOVE THE TRIAL STEM FROM THE INTRAMEDULLARY CANAL, THE SURGEON TRIED TO ENGAGE IT WITH A SECOND TRIAL OFFSET (5 MM) BUT IT BROKE AS WELL. THE SURGEON HAD TO OPEN THE TIBIAL BONE UP TO THE APEX AND THEN PERFORM A TIBIAL CERCLAGE. THERE WAS 1 HOUR AND A HALF OF DELAY IN THE SURGERY DUE TO THIS EVENT. TOTAL SURGERY TIME: ABOUT 3 HOURS.

    Patient problems: Bone Fracture(s)

    MAUDE report 18256291 ↗
  • MalfunctionNov 21, 2023Fracture; Fracture

    ON REMOVAL OF THE TRIAL ASSEMBLY COMPONENTS, FROM THE TIBIA BONE, THE ANTI-ROTATIONAL INSERT OF THE TRIAL OFFSET CAME OUT. THE TRIAL STEM REMAINED IN THE TIBIAL. IT WAS REMOVED EASILY WITH A KOCHER. NO DELAY IN THE SURGERY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18184208 ↗
  • MalfunctionAug 2, 2023Fracture; Fracture

    THE TRIAL OFFSET 3MM DETACHED FROM THE TRIAL TIBIAL BASEPLATE DURING SURGERY. THE THREADS INSIDE THE TRIAL OFFSET 3MM CAME OUT. SINCE NO THREADED PART OF THE TRIAL OFFSET REMAINED INSIDE, 40 MINUTES WERE NEEDED TO REMOVE THE TRIAL OFFSET AND THE EXTENSION STEM FROM THE TIBIAL CANAL (TOTAL SURGERY TIME OF 2 ½ HOURS). SURGERY COMPLETED SUCCESSFULLY.