Spine Instruments 03.75.10.0060Recalls & reported problems

Medacta International SA · Rod reduction device · Class 1

All models of Spine Instruments (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630971208872), received between Sep 27, 2019 and Nov 8, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING C2-T5 SPINAL FIXATION SURGERY, AT C7 LEVEL, DURING THE ROD REDUCTION PROCEDURE, THE MUST MINI REDUCTION DEVICE - SHORT AND ROD REDUCTION DEVICES WERE DISENGAGED FROM THE PEDICLE SCREW. THE REDUCTION DISTANCE TO PERFORM WAS ABOUT 3 MM. THE SURGEON PERFORMED THE ROD BENDING AGAIN AND COMPLETED THE SURGERY WITHOUT USING THE REDUCTION DEVICES. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 1 HOUR. TOTAL SURGERY TIME WAS ABOUT 5 HOURS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18091465 ↗
  • MalfunctionSep 27, 2019Difficult to Insert

    SPINAL FIXATION PERFORMED AT C2-T1. AFTER THE INSERTION OF THE ROD, THE SURGEON TRIED TO USE THE ROD REDUCTION DEVICE BECAUSE THE RODS WERE NOT INSERTED COMPLETELY INTO THE PEDICLE SCREW HEADS AT BOTH SIDE OF C5. WHEN THE SURGEON GRIPPED THE HANDLE OF THE ROD REDUCTION DEVICE, THE ROD REDUCTION DEVICE EASILY DISENGAGED FROM THE PEDICLE SCREW HEAD. THE SURGEON WAS NOT ABLE TO INSERT THE ROD BECAUSE THE SAME EVENT RECURRENTED MANY TIMES. THE SURGEON DISENGAGED THE ROD AND PERFORM THE ROD BENDING AGAIN AND FINISHED THE SURGERY WITHOUT USING THE ROD REDUCTION DEVICE. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 30 MINUTES. TOTAL SURGERY TIME WAS ABOUT 5 HOURS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9128862 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →