Skip to content

Shoulder System Instrument 04.01.10.0284Recalls & reported problems

Medacta International SA · Glenoid polyaxial screwdriver - Non-retentive · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630971211902), received between Jan 17, 2019 and May 28, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Device Fell1
  • Material Integrity Problem1
  • No Apparent Adverse Event1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 28, 2026No Apparent Adverse Event

    AFTER THE SURGERY (WOUND CLOSURE), THE X-RAYS EXAMINATION REVEALED THAT THE TIP OF THE INSTRUMENT HAD BROKEN. THE WOUND WAS REOPENED, AND THE BROKEN FRAGMENT WAS REMOVED. NO FRAGMENTS REMAIN INSIDE THE BODY. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE TOTAL SURGERY TIME WAS 3.5 HOURS, AND THE DELAY TIME WAS 1.5 HOURS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313935 ↗
  • MalfunctionNov 24, 2021Material Integrity Problem; Material Integrity Problem

    THE SURGEON HAD DIFFICULTY IMPLANTING THE GLENOSPHERE. LOOKING AT THE X-RAYS, A FRAGMENT OF THE SCREWDRIVER WAS NOTICED. THE SURGEON REMOVED THE FRAGMENT AFTER REMOVING THE IMPLANT. A NEW GLENOSPHERE WAS USED TO COMPLETE THE SURGERY. THERE WAS A 30 MINUTES DELAY, TOTAL SURGERY TIME 130 MINUTES. ALSO, THE GLENOSPHERE SCREW THREAD WAS DEFORMED.

    Patient problems: Insufficient Information

    MAUDE report 12868652 ↗
  • MalfunctionJan 17, 2019Break; Detachment of Device or Device Component; Device Fell

    DURING SURGERY, ONE PIECE OF THE TIP OF THE SCREWDRIVER WAS BROKEN. THE PIECES FELL INTO THE PATIENT AND THEN THE SURGEON RECOVERED ALL THE FRAGMENTS. THE SURGERY WAS COMPLETED WITH THE SAME SCREWDRIVER USING THREE TIPS. NO DELAY WAS REPORTED, THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Foreign Body In Patient

    MAUDE report 8256601 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates