Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07630971233805), received between Feb 20, 2025 and Sep 1, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fracture” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture6
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50543248 ↗MalfunctionMay 8, 2026Fracture
DURING THE PROCEDURE, THE DRILL BIT BROKE, RESULTING IN A SIGNIFICANT DELAY IN THE PROCEDURE (APPROXIMATELY 3 HOURS OVER 4 HOURS AND 30 MINUTES OF SURGERY) DUE TO DIFFICULTY IN REMOVING THE TIP PIECE RETRIEVED AND SURGERY COMPLETED SUCCESSFULLY. THE DRILL GUIDE WAS USED. THE PATIENT`S BONE WAS EXTREMELY SCLEROTIC, WHICH LED TO OVERHEATING OF THE DRILL BIT AND REQUIRED GREATER FORCE FOR DRILLING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25123194 ↗MalfunctionMar 25, 2025Fracture; Fracture
THE DRILL BIT FOR GLENOID POLYAXIAL SCREWS BROKE AND REMAINED IN THE GLENOID. THE SURGEON INSERTED A SMALLER SCREW TO AVOID INTERFERENCE WITH THE POLYAXIAL SCREW. YOUNG PATIENT, BONE WAS NOT SCLEROTIC; 5 MINUTES DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21687571 ↗MalfunctionMar 25, 2025Fracture; Fracture
THE DRILL BIT FOR GLENOID POLYAXIAL SCREWS BROKE AND REMAINED IN THE GLENOID. THE SURGEON INSERTED THE SCREW IN A DIFFERENT DIRECTION THAN THE DRILL. 5 MIN DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21685764 ↗MalfunctionMar 7, 2025Fracture; Fracture
THE DRILL BIT BROKE DURING THE DRILLING OF THE SUPERIOR POLYAXIAL SCREW. DRILL GUIDE WAS USED (04.01.10.0274, LOT 2159295; 04.01.10.0034, LOT 2253841). IMMEDIATELY THE OPERATION WAS STOPPED AND C-ARM DEVICE WAS APPLIED TO IDENTIFY THE POSITION OF THE BROKEN DRILL'S BIT AT THE PATIENT'S BODY. AT THE RADIOLOGICAL EXPOSURE, WAS SEEN THAT THE REMAINING PART OF THE BROKEN DRILL WAS TRAPPED IN THE CHEST, BUT NOT TO THE POINT OF CAUSING ANY VITAL PROBLEMS, ACCORDING TO THE OPINION OF A THORACIC SURGEON WHO WAS CALLED IN TO ASSESS THE SITUATION. THE OPERATION WAS CONTINUED AND COMPLETED SUCCESSFULLY (SUPERIOR AND INFERIOR SCREWS PLACED) AND WITH A VERY GOOD INTRAOPERATIVE OUTCOME IN TERMS OF JOINT'S STABILITY AND PASSIVE RANGE OF MOTION. THE SURGERY WAS EXTENDED BY ABOUT 1.5 HOURS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21547225 ↗MalfunctionFeb 20, 2025Fracture; Fracture
DURING THE SHOULDER PROSTHESIS SURGERY, THE DRILL BIT FOR POLYAXIAL SCREWS BROKE WHILE DRILLING IN THE GLENOID. A BROKEN PIECE REMAINED INSIDE THE PATIENT. DRILL GUIDE USED. SURGERY COMPLETED WITH ABOUT 10/15 MIN DELAY, TOT TIME SURGERY ABOUT 2.5 HOURS.
Patient problems: Foreign Body In Patient
MAUDE report 21418877 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page