Skip to content

Shoulder Instruments 14.00001Recalls & reported problems

Medacta International SA · Shoulder disposable pin pack · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630971259164), received between Sep 12, 2025 and Aug 4, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 4, 2026Fracture

    During the primary shoulder surgery, when the surgeon drilled his guide pin into the glenoid through the glenoid guide for his reverse total shoulder, the K-wire seemed to get stuck. As the surgeon pulled back on the guide handle, the K-wire broke off. To complete the case, the surgeon used a pin from the general instrument set. There was no delay and the surgery was completed successfully. The broken fragment was retrieved from the patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50125770 ↗
  • MalfunctionSep 12, 2025Fracture; Fracture

    WHILE REAMING THE GLENOID SURFACE, THE SURGEON CHANGED THE ANGLE WHICH PUT BENDING FORCES THROUGH THE 2.5MM GUIDE PIN AND IT THEN BROKE, LEAVING THE TIP OF THE GUIDE PIN INSIDE THE GLENOID. HE WAS UNABLE TO REMOVE THE BROKEN TIP AS IT WAS LODGED IN THE BONE, AND IT WAS LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23036831 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page