Skip to content

SALGINA 2.5mm Volar Distal Radius Plate, Narrow, Left B-111000Recalls & reported problems

Bonebridge AG · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07640187460018), received between Mar 18, 2025 and Mar 18, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Limb Fracture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTAL RADIUS FRACTURE TREATED WITH THE BONEBRIDGE SALGINA DEVICE IN SOLOTHURN, SWITZERLAND, RESULTED IN IMPLANT FAILURE SIX WEEKS POSTOPERATIVELY DUE TO EXTENSIVE SCREW DISLOCATION. THE PATIENT IS A 67-YEAR-OLD WOMAN WITHOUT A DIAGNOSED OSTEOPOROSIS, HOWEVER EXHIBITING POOR BONE QUALITY, WHICH CONTRIBUTED TO THE FAILURE. ADDITIONALLY, PATIENT NON-COMPLIANCE WAS IDENTIFIED AS A FACTOR. DETAILS FROM THE SURGICAL REPORT PROVIDED BY THE OPERATING SURGEON: THE PATIENT SUSTAINED A DISPLACED MULTIFRAGMENTARY INTRA-ARTICULAR FRACTURE OF THE LEFT RADIUS ON (B)(6) 2024. OPEN REDUCTION AND PALMAR PLATE OSTEOSYNTHESIS OF THE DISTAL LEFT RADIUS WAS PERFORMED ON (B)(6) 2025, USING THE SALGINA. DURING THE PROCEDURE, THE BONE WAS OBSERVED TO BE VERY SOFT; HOWEVER, INTRAOPERATIVE IMAGING CONFIRMED PROPER FRACTURE REDUCTION AND CORRECT IMPLANT POSITIONING. POSTOPERATIVELY, THE WRIST WAS IMMOBILIZED WITH A SPLINT, WHICH WAS LATER REPLACED WITH A FOREARM CAST FOR SIX WEEKS. WEIGHT-BEARING AND RESISTANCE EXERCISES WERE RESTRICTED DURING THIS PERIOD. THE SIX-WEEK POSTOPERATIVE FOLLOW-UP ON (B)(6) 2025, REVEALED SCREW LOOSENING AND FULL FRACTURE DISPLACEMENT, ACCOMPANIED BY A DORSAL TILT OF APPROXIMAT

    Patient problems: Limb Fracture

    MAUDE report 21635488 ↗

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates